Maintenance

Replacing Capillaries on a Standard Flow Cell

ag-textnode-version-id: 3358984715
ag-textnode-base-id: 3358984715

Replacing Capillaries on a Standard Flow Cell

For bio-inert modules use bio-inert parts only!
When
Tools required
Qty.
1
1
1
1
Parts required
Qty.
1
Preparations
All descriptions in this procedure are based on the default orientation of the cell
NOTE
(as it is manufactured). The heat exchanger/capillary and the cell body can be
fixed mirror symmetrically to have both capillaries routed to the bottom or to the
top (depending on the routing of the capillaries to the column).
The fittings at the flow cell body are special types for low dead volumes and not
NOTE
compatible with other fittings.
When retightening the fittings, make sure that they are carefully tightened
(handtight plus 1/4 turn with a wrench). Otherwise damage of the flow cell body
or blockage may result.
Diode Array Detector WR and Multiple Wavelength Detector User Manual
If the capillary is blocked
p/n
5043-0915
p/n
Turn the lamp(s) off.
Open the doors of the module.
Remove the flow cell, see
Description
Wrench, 1/4 inch
Fitting mounting tool
Wrench, 4 mm
Screwdriver, Pozidriv #1 PT3
Description
For parts see
Standard Flow Cell
Remove and Install a Flow Cell
on page 236
on page 188.
201