Error Information
Detector Error Messages
ag-textnode-base-id: 3284016651
ag-textnode-version-id: 9007202538757643
UV Lamp: No Voltage
Error ID: 7451
The lamp anode voltage is missing. The processor continually monitors the
anode voltage across the lamp during operation. If the anode voltage falls below
the lower limit, the error message is generated.
Probable cause
1
2
3
Diode Array Detector WR and Multiple Wavelength Detector User Manual
Defective UV lamp or non-
Agilent lamp.
Defective power supply.
Defective mainboard.
Suggested actions
• Exchange the UV lamp.
• Please contact your Agilent service representative.
• Please contact your Agilent service representative.