Stryker Arise 1000EX Operation & Maintenance Manual

Stryker Arise 1000EX Operation & Maintenance Manual

Low air loss therapy mattress

Advertisement

Quick Links

Arise
1000EX
®
Low Air Loss
Therapy Mattress
Model 2236
Operations/Maintenance Manual
For Parts or Technical Assistance:
USA: 1-800-327-0770 (option 2)
2008/11
2236-009-001 REV A
www.stryker.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Arise 1000EX and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Stryker Arise 1000EX

  • Page 1 Arise 1000EX ® Low Air Loss Therapy Mattress Model 2236 Operations/Maintenance Manual For Parts or Technical Assistance: USA: 1-800-327-0770 (option 2) 2008/11 2236-009-001 REV A www.stryker.com...
  • Page 3: Table Of Contents

    Arise 1000EX Assembly - 2236-000-001 ........
  • Page 4 Service ................27 2236-009-001 REV A www.stryker.com...
  • Page 5: Symbols And Definitions

    This provides special information to make maintenance easier or important instructions clearer. Return To Table of Contents www.stryker.com 2236-009-001 REV A...
  • Page 6: Introduction

    (Fowler Boost). Additionally, low friction and shear materials, together with average interface pressures below capillary closure levels (32mm of Mercury), means that the Arise® meets the comfort and clinical requirements. Return To Table of Contents 2236-009-001 REV A www.stryker.com...
  • Page 7: Specifications

    Ambient Temperature 100% Relative Humidity (Non-Condensing) Stryker reserves the right to change specifications without notice. Specifications listed are approximate and may vary slightly from unit to unit or by power supply fluctuations. Return To Table of Contents www.stryker.com 2236-009-001 REV A...
  • Page 8: Therapy Mattress System Illustration

    ThErAPy MATTrESS SySTEM iLLuSTrATiOn Therapy Control unit Therapy Mattress System ThErAPy COnTrOL uniT iLLuSTrATiOn Return To Table of Contents 2236-009-001 REV A www.stryker.com...
  • Page 9: Product Features

    Five-inch deep therapy cells are constructed of highly durable, polyurethane coated nylon to provide adequate support and prevent bottoming out for most patients within the weight limit. • Maximum weight capacity: 1000 lbs.* Return To Table of Contents www.stryker.com 2236-009-001 REV A...
  • Page 10: Indications For Use

    Breathable, moisture vapor permeable fabric allows air to circulate beneath the pad and wicks away excess moisture. This keeps your patient dry and helps to prevent skin maceration. • Two-way stretch therapy pad is designed for optimal comfort, moisture vapor transfer, stain resistance and ease of laundering. Return To Table of Contents 2236-009-001 REV A www.stryker.com...
  • Page 11: Key Areas In The Treatment Of Compromised Skin

    When a patient’s skin rubs against another surface, the result is friction. The top surface of the Arise® therapy pad is constructed from a very smooth nylon fabric with low friction and shear properties to protect the patient’s skin from these damaging forces. Return To Table of Contents www.stryker.com 2236-009-001 REV A...
  • Page 12: Summary Of Safety Precautions

    • Do not attempt to service the control unit. Please call Stryker at 1-800-327-0770 for any service requests. • The therapy pad (top cover) of this product is not air permeable and may present a suffocation risk. It is the responsibility of the caregiver to ensure that the patient can use this product safely.
  • Page 13: Bed System Entrapment Information

    For a higher level of assurance, an electrician should be requested to thoroughly test the electrical system with more reliable equipment. If repair or replacement of the cord or plug is necessary, please contact Stryker at 1-800-327-0770. figure 2 Return To Table of Contents www.stryker.com...
  • Page 14: Setup Procedures

    Allow three to five minutes for full inflation. Place the patient on the mattress. Mattress can be inflated with patient on it, but will take longer, depending on patient weight/size. note: Keep the control unit on while the patient is on the mattress. Return To Table of Contents 2236-009-001 REV A www.stryker.com...
  • Page 15 Inflate the extension foot cells by turning the lower left knob on the mattress control panel (see Figure 5) up. After 2-5 minutes the cells will inflate. Leave the knob in this position to maintain inflation. figure 5 Return To Table of Contents www.stryker.com 2236-009-001 REV A...
  • Page 16: Operation Guide

    The unit is designed to lock out all the adjustment controls after the patient has been positioned correctly. In approximately five minutes after the last button push, the power on light begins to flash indicating Lockout is enabled. Return To Table of Contents 2236-009-001 REV A www.stryker.com...
  • Page 17: Cpr

    Note: Side bolsters are meant to be used with side rails and to provide a documentable and functional intervention for the risk management issues of falls and entrapment but in no way guarantees the prevention of falls or entrapment occurrences. Return To Table of Contents www.stryker.com 2236-009-001 REV A...
  • Page 18: Nursing Procedures

    To minimize the risks of falls or injury the patient surface should always be in the lowest practical position when the patient is unattended. Make sure areas under and around the frame are clear of objects, persons and parts of body before adjusting height. Return To Table of Contents 2236-009-001 REV A www.stryker.com...
  • Page 19: Cleaning, Storage And Care

    Allow all components to air dry. Wrap mattress in plastic and return to storage area. • Thoroughly wipe down outside of control unit and allow to air dry. Cover with plastic and return to storage area • Remove gloves and dispose; wash hands. Return To Table of Contents www.stryker.com 2236-009-001 REV A...
  • Page 20: Therapy Pad

    Support Surface: Gently roll up the support surface, expelling any residual air, for temporary storage. The mattress should be wrapped in plastic and/or a clean bag for storage. Return To Table of Contents 2236-009-001 REV A www.stryker.com...
  • Page 21: Preventative Maintenance

    _____ Plastic on the control unit assembly is not damaged. _____ Mattress holds air and all straps are intact. _____ Inspect all product labeling for signs of degradation. Unit Serial Number: Completed by: _______________________________________ Date: _________________ Return To Table of Contents www.stryker.com 2236-009-001 REV A...
  • Page 22: Troubleshooting

    Ensure that the power is button. Unable to change therapy mode Make sure the Lockout function is disabled. To disable, press or adjust comfort control. the up and Down comfort adjust arrows simultaneously. Return To Table of Contents 2236-009-001 REV A www.stryker.com...
  • Page 23: Quick Reference Replacement Parts List

    Quick reference replacement Parts List note The parts and accessories listed on this page are all currently available for purchase. Please call Stryker Customer Service (800)-327-0770. Arise® 1000EX Power unit: 2236-100-002 Part name Part view SyK Part number Blower assembly...
  • Page 24 Air Cell, Cari Rail & Side Extension 2236-100-009 Air Cell, Foot Extension (48”) 2236-100-029 Air Cell, Main Support (5x36”) 2236-100-028 Back Raise Sensor Assembly 2236-100-032 Control Panel, Arise 1000 EX 2236-100-013 Return To Table of Contents 2236-009-001 REV A www.stryker.com...
  • Page 25 2236-100-007 External Hose Cover, Bariatric 2236-100-024 Foam Safety Mattress, (3”) 2236-100-012 Hose Clamp (Pkg 25) 2236-100-034 Loop Strap 2236-100-011 Mattress Tub, Arise 1000 EX 2236-100-025 Therapy Cover, Arise 1000 EX 2236-100-006 Return To Table of Contents www.stryker.com 2236-009-001 REV A...
  • Page 26: Arise 1000Ex Assembly - 2236-000-001

    Arise 1000EX Assembly - 2236-000-001 For Reference Only: 55-0059 Rev-00 item Part no. Part name Qty. 2236-100-001 Arise® 1000EX Power Unit 2236-100-002 Arise® 1000EX Mattress Return To Table of Contents 2236-009-001 REV A www.stryker.com...
  • Page 27 For further service needs please call Stryker at 1-800-327−0770. * Does not include maintenance due to abuse or for any disposable items. Stryker reserves the right to change options without notice. Return To Table of Contents www.stryker.com...
  • Page 29 UNITED STATES UNITED STATES Stryker Medical Stryker Medical 3800 E. Centre Ave., 3800 E. Centre Ave., Portage, Michigan USA Portage, Michigan USA 49002 49002 CANADA CANADA Stryker Canada Stryker Canada 45 Innovation Drive 45 Innovation Drive Hamilton, Ontario Canada Hamilton, Ontario Canada...