Device Clear On Disconnect - Keysight E6680A User & Programmers Manual

Signal analyzer iq analyzer mode
Hide thumbs Also See for E6680A:
Table of Contents

Advertisement

See the example below, where the invalid command
Normal response to
SCPI> SENS:BOGUS
SCPI> SYST:ERR?
-113,"Undefined header"
After turning on Verbose SCPI:
SCPI> SYST:BOGUS
SCPI> SYST:ERR?
-113,"Undefined header;SYST:BOGUS<Err>"
:SYSTem:ERRor:VERBose OFF | ON | 0 | 1
Remote Command
:SYSTem:ERRor:VERBose?
:SYST:ERR:VERB ON
Example
Preset
Unaffected by Preset, but set to
State Saved
No
OFF|ON
Range

4.2.2.6 Device Clear on Disconnect

When using HiSLIP (High Speed LAN Instrument Protocol), Telnet, or Sockets, a
communication session with the instrument is opened when you connect, and
closed when you disconnect. This differs from other connections such as GPIB, USB
and VXI-11 connections, which are never actually closed but stay open as long as
the instrument is running.
When a session is closed, a Device Clear function is generated, which affects the
entire instrument, not just the current connection. Thus, when using HiSLIP, Telnet,
or Sockets, unexpected Device Clears may occur, which can disrupt measurements
in ways that GPIB and VXI-11 "sessions" do not.
Device Clear on Disconnect enables these auto-generated Device Clears for Telnet,
Socket, and HiSLIP sessions. For backwards compatibility, they are not generated
unless you explicitly enable them.
There is no change in VXI-11, USB, or GPIB session behavior. These sessions do not
close when you disconnect, have never generated Device Clear events, and still do
not generate Device Clear events, regardless of the setting of this switch.
:SYSTem:COMMunicate:LAN:SCPI:EOSession:DCLEar:ENABle 0 | 1 | ON | OFF
Remote
Command
:SYSTem:COMMunicate:LAN:SCPI:EOSession:DCLEar:ENABle?
:SYST:COMM:LAN:SCPI:EOS:DCLE:ENAB ON
Example
Preset
Unaffected by Preset, but set to
IQ Analyzer Mode User's & Programmer's Reference
:SYST:ERR?
(using the Telnet window):
OFF
by Restore Defaults
OFF
by Restore Defaults
:SENS:BOGUS
>"Misc" on page 482
>"Misc" on page 482
4  System
4.2  I/O Config
is sent:
439

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the E6680A and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents