Instructions for use SMARTmatic S80 K – 1.011.6771 Table of contents Table of contents 1 User instructions....................... 2 Safety ..........................2.1 Infection hazard......................7 2.2 Improper use ....................... 7 2.3 Technical condition......................8 2.4 Accessories and combination with other equipment ............8 2.5 Qualification of personnel....................
SMARTmatic is a registered trademark of KaVo Dental GmbH. All other trademarks are property of their respective owners. KaVo Original Factory Repair In the event of a repair, please ship your product to the KaVo Original Factory Repair using www.kavobox.com. KaVo Technical service...
Page 5
Instructions for use SMARTmatic S80 K – 1.011.6771 1 User instructions Serial number UDI symbol Manufacturer Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code CE mark for medical devices EAC conformity mark (Eurasian Conformity) UA Mark of conformity UA.TR.099...
Page 6
Instructions for use SMARTmatic S80 K – 1.011.6771 1 User instructions WARNING In cases which – if not prevented – can lead to death or severe injury. CAUTION In cases which – if not prevented – can lead to minor or moderate injury.
Instructions for use SMARTmatic S80 K – 1.011.6771 2 Safety | 2.1 Infection hazard 2 Safety NOTE All serious events occurring in relation to the product must be reported to the manufacturer and the competent authority of the member state, in which the user and/or patient resides.
2 Safety | 2.3 Technical condition 2.3 Technical condition A damaged product or damaged or NOT KaVo original components could injure patients, users or third parties. 4 Only operate devices or components if they show no signs of damage on the outside.
Define the service interval depending on the frequency of use. As a result of the use of NON-KaVo original spare parts during the repair, parts such as covers may become undone and injure the patient, user or other people.
Instructions for use SMARTmatic S80 K – 1.011.6771 3 Product description | 3.1 Intended use 3 Product description SMARTmatic S80 K (Mat. no. 1.011.6771) 3.1 Intended use Indications for use: This medical device is: ▪ Intended for dental treatment only. All other types of use of or modifications to the product are not permitted and can be hazardous.
Instructions for use SMARTmatic S80 K – 1.011.6771 3 Product description | 3.3 Transportation and storage conditions 3.3 Transportation and storage conditions CAUTION Startup after refrigerated storage. Malfunction. 4 Prior to startup, strongly refrigerated products must be allowed to warm up to a temperature of 20 °C to 25 °C (68 °F to 77 °F).
Dispose of and recycle the packaging appropriately in accordance with current packaging law, employing waste management / recycling companies. Comply with the comprehensive return system. KaVo has had its packaging licensed for this purpose. Please comply with the regional public waste-disposal system.
Instructions for use SMARTmatic S80 K – 1.011.6771 5 Operation | 5.1 Attaching the medical device 5 Operation WARNING Rotating dental tool. Cuts, infection and burn injury. 4 Never push the press-button while the dental tool is rotating. 4 Do not touch the dental tool while it is rotating.
CAUTION Dental tool with damaged, worn or deformed shafts. A damaged product or damaged or NOT KaVo original components can injure patients, users or third parties. Injury hazard, dental tool may fall out during treatment. 4 Never use dental tools with damaged, worn or deformed shafts.
Instructions for use SMARTmatic S80 K – 1.011.6771 5 Operation | 5.4 Removing the dental bur CAUTION Dental tool with damaged, worn or deformed shafts. Material damage to the chuck system, dental tool is difficult or impossible to re- move from the chuck system.
Malfunction or complete failure. 4 Do not use Vaseline or other grease or oil. 4 Spray a lint-free cloth with KaVo Spray and apply it to the O-rings on the coupling. 4 Press the O-ring between your fingers to form a loop.
Page 17
Instructions for use SMARTmatic S80 K – 1.011.6771 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting 4 Spray a lint-free cloth with KaVo Spray, apply it to the new O-rings and in- sert them in the recesses. 17 / 28...
Malfunction and material damage. 4 Reprocess in a washer disinfector or by hand only. KaVo recommends the following products based on the compatibility of the ma- terials. The microbiological efficacy must be ensured by the disinfectant manu- facturer and proven by an expert opinion.
Instructions for use SMARTmatic S80 K – 1.011.6771 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.2 Manual reprocessing 7.2 Manual reprocessing WARNING Incomplete disinfection. Patients, users or third parties could be infected by contaminated medical devices.
7.2.3 Manual drying Use KaVo DRYspray for subsequent drying of the air, water and gear unit ducts. 4 Cover the medical device with the KaVo Cleanpac bag, and place it on the corresponding care adapter. 4 Hold the can vertically.
NOTE Please comply with the instructions for use of the washer disinfector. 4 In order not to affect the KaVo medical device, make sure that the product is dry on the inside and outside after completion of the cycle. 4 Remove any residual liquids with KaVoDRYspray.
30 minutes of operation. 4 Remove the dental tool from the medical device. 4 Cover the medical device with the KaVo Cleanpac bag, and place it on the corresponding care adapter. 4 Press the spray key for 1 to 2 seconds.
Instructions for use SMARTmatic S80 K – 1.011.6771 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.5 Packaging NOTE End the chuck service mode. Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A, close the front flap and start the service procedure.
Page 24
Instructions for use SMARTmatic S80 K – 1.011.6771 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.7 Storage NOTE Comply with the expiry date of the sterilised items. 24 / 28...
Instructions for use SMARTmatic S80 K – 1.011.6771 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat.No. INTRA Instrument stand 3.005.5204 Cleanpac 10 units 0.411.9691 Cellulose pad 100 units 0.411.9862 KaVo CLEANspray 2110 P 1.007.0579...