Sensidyne Gilian LFS-113 Operation Manual

Sensidyne Gilian LFS-113 Operation Manual

Low flow air sampling pump
Hide thumbs Also See for Gilian LFS-113:

Advertisement

Quick Links

LFS-113 LOW FLOW
AIR SAMPLING PUMP
OPERATION MANUAL
1000 112th Circle N, Suite 100 • St. Petersburg, Florida 33716 USA
(800) 451-9444 • (727) 530-3602 • (727) 539-0550 [FAX] • www.sensidyne.com
FLOW
ADJUST
LFS-113DC
LOW FLOW SAMPLER
TIME (MIN)
WARNING
Possible static hazard. Do not rub with dry cloth.
Substitution of components may impair intrinsic safety.
Use only specified battery pack. Do not charge in
hazardous locations. Use only with specified charger.
Max. Charging Voltage Um: 21 Vrms • Max. Charging Current: 250 mA
LFS-113D & LFS-113DC are intrinsically safe for use in Zone 1 & 2 hazardous locations.
0518
II 2 G
SIRA 03ATEX2137
FM 3610
EEx ib IIC T4
Manufactured by Sensidyne, LP • St. Petersburg, Florida 33716 USA
800-451-9444 • 727-530-3602 • www.sensidyne.com
REF 360-0041-01 (G)
B
ON
F
OFF
Ex ib IIC T4
AEx ib IIC T4
2003 1469304X

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Gilian LFS-113 and is the answer not in the manual?

Questions and answers

Summary of Contents for Sensidyne Gilian LFS-113

  • Page 1 Manufactured by Sensidyne, LP • St. Petersburg, Florida 33716 USA 800-451-9444 • 727-530-3602 • www.sensidyne.com 1000 112th Circle N, Suite 100 • St. Petersburg, Florida 33716 USA (800) 451-9444 • (727) 530-3602 • (727) 539-0550 [FAX] • www.sensidyne.com REF 360-0041-01 (G)
  • Page 3: Notices

    PROPRIETARY NOTICE This manual was prepared exclusively for the owner of the Sensidyne LFS-113 Low Flow Air Sampling Pump. The mate- rial within this manual is proprietary information and is to be used only to understand, operate, and service the instrument.
  • Page 4: Table Of Contents

    TABLE OF CONTENTS • PREFACE • Notices ........................... 2 • WARNINGS ..........................4 • Certifications .......................... 5 SECTION ONE: INTRODUCTION • Components .......................... 6 SECTION TWO: PUMP OPERATION Equipment Set-Up ......................9 Field Calibration/Flow Verification ................10 Taking A Sample ......................13 SECTION THREE: MAINTENANCE Battery Maintenance .......................
  • Page 5: Warnings

    DO NOT operate this product should it malfunction, require repair, or have a cracked or broken case. DO NOT repair or modify, except as specified in Operation Manual. No user servicable parts located within case. Service to be performed by Sensidyne Authorized Service Departments only.
  • Page 6: Certifications

    II = Equipment Group FM 3610 EEx ib IIC T4 2 = Equipment Category CSA marking Manufactured by Sensidyne, LP • St. Petersburg, Florida 33716 USA G = Hazardous Gases, vapors or mists 800-451-9444 • 727-530-3602 • www.sensidyne.com EC-Type examination certification number...
  • Page 7: Components

    SECTION ONE INTRODUCTION • COMPONENTS • See Figures 1.1 and 1.2 (1) Inlet Boss (9) Discharge (Bag Sampling) Boss The Inlet Boss is located on the filter housing and This is an accessory which, when installed into provides a built-in means of attaching tubing for the Outlet Port, provides a means for filling air suction sampling.
  • Page 8: Lfs-113 Air Sampling Pump: Front View

    0518 II 2 G 2003 1469304X SIRA 03ATEX2137 FM 3610 EEx ib IIC T4 Manufactured by Sensidyne, LP • St. Petersburg, Florida 33716 USA 800-451-9444 • 727-530-3602 • www.sensidyne.com Figure 1.1 LFS-113 Air Sampling Pump: Front View REF 360-0041-01 (G)
  • Page 9: Lfs-113 Air Sampling Pump: Rear View

    LFS-113 LOW FLOW AIR SAMPLING PUMP Figure 1.2 LFS-113 Air Sampling Pump: Rear View REF 360-0041-01 (G)
  • Page 10: Equipment Set-Up

    II 2 G AEx ib IIC T4 2003 1469304X SIRA 03ATEX2137 FM 3610 EEx ib IIC T4 Manufactured by Sensidyne, LP • St. Petersburg, Florida 33716 USA 800-451-9444 • 727-530-3602 • www.sensidyne.com Figure 2.1 Field Calibration Equipment Set-Up REF 360-0041-01 (G)
  • Page 11: Field Calibration/Flow Verification

    LFS-113 LOW FLOW AIR SAMPLING PUMP 2.2 Field Calibration/Flow Verification Field calibration (flow rate verification) must be per- 5) Turn on the pump using a pointed instrument formed before sampling and when setting the flow such as ball point pen [5]. rate.
  • Page 12: Field Calibration: Constant Flow

    II 2 G AEx ib IIC T4 2003 1469304X SIRA 03ATEX2137 FM 3610 EEx ib IIC T4 Manufactured by Sensidyne, LP • St. Petersburg, Florida 33716 USA 800-451-9444 • 727-530-3602 • www.sensidyne.com Figure 2.2 Field Calibration: Constant Flow REF 360-0041-01 (G)
  • Page 13: Field Calibration: Multi-Flow

    AEx ib IIC T4 0518 II 2 G 2003 1469304X SIRA 03ATEX2137 FM 3610 EEx ib IIC T4 Manufactured by Sensidyne, LP • St. Petersburg, Florida 33716 USA 800-451-9444 • 727-530-3602 • www.sensidyne.com Figure 2.3 Field Calibration: Multi-Flow REF 360-0041-01 (G)
  • Page 14: Taking A Sample

    EEx ib IIC T4 EEx ib IIC T4 FLOW ADJUST Manufactured by Sensidyne, LP • St. Petersburg, Florida 33716 USA Manufactured by Sensidyne, LP • St. Petersburg, Florida 33716 USA 800-451-9444 • 727-530-3602 • www.sensidyne.com LFS-113DC 800-451-9444 • 727-530-3602 • www.sensidyne.com...
  • Page 15: Battery Maintenance

    SECTION THREE MAINTENANCE 3.1 Battery Maintenance • Chargers NOTE Single Station Charger Do not charge battery pack while A dual rate charger that can be switched from in an explosive atmosphere. constant-rate charge to trickle charge. The LFS-113 pump uses rechargeable Nickel-Metal- LFATR Multi-Station Charger Hydride batteries that must be fully charged and properly A dual rate charger offering five-station timed...
  • Page 16: Appendix A: Parts List

    APPENDIX A PARTS LIST Accessories Part Number Description 200504 ......Tubing, 1/8” ID x 1/16” W (10 ft) 200505 ......Tubing, 1/8” ID x 1/16” W (3 ft) 800565-4 ..... Diagnostic Panel with Carrying Case (2–50cc, 20–200cc, 50–800cc) 800400 ......Carrying Case (18” x 13” x 7”) 800093 ......
  • Page 17: Appendix B: Specifications

    APPENDIX B SPECIFICATIONS Additional Features ........Flow Fault indication LED Batt check LED Belt clip Dual filtration system Sorbent tube end breaker External flow adjust Options ............Elapsed timer clock module (DC models only): LCD display automatic instant-fault shutdown function RFI shielding Dimensions ...........
  • Page 18: Appendix F: Service

    APPENDIX C SERVICE Domestic Service European Service Sensidyne, LP. Goffin Meyvis 1000 112th Circle N, Suite 100 Analytical and Medical Systems B.V. St. Petersburg, Florida 33716 USA Deliveries: 800-451-9444 727-530-3602 Ecustraat II 4879 NP Etten Leur 727-539-0550 [Main fax] the Netherlands...
  • Page 20 Schauenburg, Electronic Technologies GmbH St. Petersburg, Florida 33716 Weseler Str. 35 • 45478 Mulheim-Ruhr Germany 800-451-9444 • 727-530-3602 • 727-539-0550 [fax] Tel. +49 (0) 208 9 99 10 • Fax. +49 (0) 208 5 41 10 www.sensidyne.com • info@sensidyne.com www.schauenburg.com • international@schauenburg.com...