Sensidyne Gilian 5000 Basic Manual

Sensidyne Gilian 5000 Basic Manual

Hide thumbs Also See for Gilian 5000:

Advertisement

Quick Links

Gilian 5000
This manual covers specifications, warnings and basic operation. For compete information,
including detailed operation, options, and other details see the Operation Manual (PN 360-
0103-01). The warnings, safe operation, installation and maintenance instructions in that
manual should be adhered to at all times.
Cautions:
Pump is intrinsically safe for use in areas designated in the attached specifications listed on
page 4. Do not replace or charge batteries in hazardous area. Charge batteries completely
before each use. No special discharge or battery conditioning required. Use only specified
charger, and charge only within specified temperature range.
Operation Guide
Keypad Operation
Power/Enter: Press momentarily to turn on.
While on, press for five seconds to turn off.
Also used to confirm entries and adjustments.
Set/Cal: Used to select flow rate set (FLO) or
Calibration (CAL)
▲/Clear: Used in set modes to adjust
displayed value upward. Also used to reset
accumulated run data before initiating new
sample when held down for 15 seconds.
▼/Run/Stop: Used in set modes to adjust
displayed value downward. Also used to start
and stop sampling when held down for five
seconds.
Turning Pump power ON and OFF
On: Press Power/Enter key momentarily to
turn power on. Display turns on all
segments and indicators briefly, shows
the software revision, shows the number
of hours since last calibration, then
enters the Ready mode.
Off: While the pump is in Ready mode (not
running), press and hold the
Power/Enter key for several seconds, until the display shows "OFF". After a few
seconds, the pump will turn power off.
16333 Bay Vista Dr. • Clearwater, FL 33760 USA
(800) 451-9444 • (727) 530-3602
(727) 539-0550 [FAX]
www.sensidyne.com
Basic Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Gilian 5000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Sensidyne Gilian 5000

  • Page 1 16333 Bay Vista Dr. • Clearwater, FL 33760 USA (800) 451-9444 • (727) 530-3602 (727) 539-0550 [FAX] www.sensidyne.com Gilian 5000 Basic Manual This manual covers specifications, warnings and basic operation. For compete information, including detailed operation, options, and other details see the Operation Manual (PN 360- 0103-01).
  • Page 2 Gilian 5000 Air Sampling Pump Setting the Flow Rate: 1 While the pump is in Ready mode, press SET button once. “FLO” is displayed. 2 Press ENTER button to begin setting the flow rate. 3 Press and hold ▲ button to increase flow rate set point or ▼ button to decrease flow rate set point.
  • Page 3 Maintenance Battery The Gilian 5000 pump uses rechargeable Nickel-Metal-Hydride batteries that must be fully charged and properly maintained for maximum run time. The battery pack has a charge time under 4 hours. Make certain charger plug is fully inserted into jack on battery pack.
  • Page 4: Specifications

    Battery Pack ...........Removable, Sealed, NiMH (7.2v) Run Time ..........8 hours at maximum load (4000cc/min @ 30”) Charge Time ...........< 4 hours (Use only Sensidyne PN 298-0013-01) Intrinsic Safety ........FM 07ATEX0018X US/Canada (Hazardous....Class I, II, III Div 1, Groups A, B, C, D; E, F, G T4 Area Certification) ......Class I, Zone 0, Group IIC T4...

Table of Contents