Omron ZS-L User Manual page 93

Zs-l series smart sensor
Hide thumbs Also See for ZS-L:
Table of Contents

Advertisement

FUN mode-[I/O SET]-[ANALOG]-[FOCUS]-[ON]
1.
Set the output value (voltage or current value) of
point 1.
The output value is displayed on the main display.
2.
Press the SET Key to apply the setting.
3.
Set the measurement value for point 1.
The measurement value is displayed on the sub-dis-
play.
4.
Press the SET Key to apply the setting.
5.
Set point 2 in the same way as point 1.
If the points are not set correctly, check the following:
• Is the measured value set on the sub-display within the measuring range (with scaling and calcula-
tion settings reflected if set)?
• Are the first and second measured points separated by at least 1% of the rated measuring range?
• Are the current (or voltage) values for the two points the same?
Section 4
Linear Input/Output
INPUT POINT 1
CHANGE SET:OK
INPUT POINT 1
DIG
VAL SET:OK
INPUT POINT 2
DIG
VAL SET:OK
User's Manual
ZS-L
4-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zs-ld20tZs-ld40tZs-ld50Zs-ld80Zs-ld200

Table of Contents