USER MANUAL FOR TELESCOPIC
A688-00
A689-00
.
2023_09_23
4 Le Pas du Château
85670 SAINT PAUL MONT PENIT
TEL : +33 (0)2 51 98 55 64
FAX : +33 (0)2 51 98 59 07
Email :
info@medicatlantic.fr
Website: http://www.winncare.fr
ALUMINIUM BARRIERS
580314 English

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the A688-00 and is the answer not in the manual?

Questions and answers

Summary of Contents for Winncare A688-00

  • Page 1 4 Le Pas du Château 85670 SAINT PAUL MONT PENIT TEL : +33 (0)2 51 98 55 64 FAX : +33 (0)2 51 98 59 07 Email : info@medicatlantic.fr Website: http://www.winncare.fr USER MANUAL FOR TELESCOPIC ALUMINIUM BARRIERS A688-00 A689-00 2023_09_23...
  • Page 2: Table Of Contents

    Consequently, compliance with the conditions of use recommended by WINNCARE and the use of original panels and accessories condition the maintenance of the warranty clauses of the contract and ensure the safe use of the medical bed and its accessories.
  • Page 3: Installation Conditions

    3.2. Assembly WINNCARE telescopic barriers are designed to be compatible with the panels provided for this purpose (Stylvia, Novida, etc.). These must be combined with mattresses whose technical characteristics are indicated in the bed's instructions. For assembly : 1-Insert the two clamps on the bed base, then insert the TRCC screws and spacers.
  • Page 4: Conditions Of Use

    Location labels are affixed to each rail, indicating their possible positions on the bed. 4. CONDITIONS OF USE 4.1. Function The barriers are designed to prevent patients from falling out of bed while they are sleeping and being transported BUT they are not designed to prevent a patient from voluntarily getting out of bed. Many accidents occur when patients try to get out of bed despite the barriers being in place.
  • Page 5: Precautions For Use

    2-In the up position :  Lift the handrail with both hands, pressing the two release fingers all the way up and then releasing the fingers before the top stop.  Check that it is engaged by trying to lower it without touching the release finger. ➢...
  • Page 6: Residual Risks

    (if the patient's condition so requires: confused, demented or physically weak people). If you need assistance with installation, operation or maintenance, or if you want to report unexpected events or malfunctions, call your supplier or Winncare. 4.4. Residual risks...
  • Page 7: Cleaning

    5.3. Cleaning High-pressure cleaning is prohibited. Clean the structure using soapy water and a soft cloth, then rinse and dry. Do not use aggressive or scouring products such as detergents, powders, solvents or bleach. 5.4. Disinfection Isolate the barrier in a disinfection room equipped with a particle filtration system and a drain for washing floors and walls after disinfection.

This manual is also suitable for:

A689-00