1. TRANSPORT AND STORAGE 2. BED ENVIRONMENT CONDITIONS 3. GENERAL USE 3.1. P REREQUISITES BEFORE USE AND INSTRUCTIONS FOR USE 3.1.1. T RAINING AND QUALIFICATIONS OF THE PERSON USING THE DEVICE 3.1.2. I NDICATIONS 3.1.3. C ONTRAINDICATIONS 3.2. C LINICAL BENEFIT DEVICE PERFORMANCE ACTION MECHANISM DEPENDING ON THE FUNCTIONS OPTIONS OR ACCESSORIES USED...
Page 3
7.1.3. E MERGENCY RELEASE OF THE BACK REST ARDIO ULMONARY ESUSCITATION 7.2. L EG REST 7.2.1. L (C/) EG REST WITH MANUAL CRANK 7.2.2. L (P/) EG REST WITH ELECTRIC FOLDING 8. INSTALLATION OF ACCESSORIES 8.1. B OARDS 8.2. M ETAL SIDE RAILS 8.3.
Dear Sir/Madam, You have acquired a WINNCARE medical bed equipped with its accessories, and we thank you for your custom. Our class I medical devices are designed and manufactured in accordance with Annex VIII of Regulation (EU) 2017/745. They are tested in conformity with standard IEC 60601-2-52 (2010) in their commercial configurations, including the boards and accessories that we manufacture, so as to ensure you maximum safety and performance.
The user and staff must be trained and aware of the risks associated with using the bed. He must not allow it by children and be vigilant when used by confused or disoriented people. 3.1.2. Indications Suitable for patients with temporary or permanent decreased independence due to their illness or disability. 3.1.3.
Page 6
The wired control must be hooked to the headboard when not in use. In the case of the use of infrared remote control(s), WINNCARE allows the establishment of a single bed in the same room (or in a close environment) or a second bed only if the infrared options of 2 beds concerned are different (I and I1).
Plug in the power lead, checking that the mains comply with the standards in force and that it is suitable for the supply box voltage. Also ensure that the power lead, the remote control lead as well as the cables of possible other devices are positioned correctly to prevent any risks of getting caught between the moving parts of the bed.
3.4.2. Electromagnetic compatibility The bed is an electromedical device requiring special precautions with regard to electromagnetic compatibility. The device must be installed and put into service according to the EMC information provided in this manual. The use of accessories, transducers, and cables other than those specified or supplied by the manufacturer may cause increased electromagnetic emissions or a decrease in the immunity of the device and may cause improper operation.
5. COMPATIBLE ACCESSORIES Only accessories and boards supplied by WINNCARE guarantee you a compatibility with our products. Incompatible accessories and boards can pose RISKS...
6. USE 6.1. Purpose of device ➢ The device is intended to compensate for a handicap and as a preventive measure. ➢ Depending on their configuration, Aerys 120 beds are intended for home use HAD – MAD (application environment 4) (only with option u), EPHAD and HPA (application environment 3). ➢...
Raise the Aerys bed to a minimum height of 30 cm when using a patient lift or dining table. WINNCARE recommends the use of a patient lift or a dining table bases with less than 13.5cm high. 67 cm 39.5 cm 55.5 cm...
6.4.2. Weight Normal load in use: 315 kg (Patient 270 kg, Mattress 30kg. Accessories 15kg) Width 120 cm Lit Aerys (P/) 101 kg Bed, version electric folding leg rest (without boards) 6.4.3. Noise The measurement of the maximum audible sound power in accordance with ISO 3746 is 53 dB(A). 6.4.4.
6.4.5. Protection level against dust and liquid penetration number (decade) number (unit) Index Protection against solids Protection against water intrusion No protection. No protection. Protected against solid bodies greater than Protected against vertical drops of water drops. 50 mm. Protected against solid bodies greater than Protected against falling drops of water up to 15°...
6.6. Remote Control AERYS 6.6.1. Custom Variable height position Possibility on all remote control of Aerys bed (except IR, use the wired caregiver case) to store a stop on a custom position of the variable height (factory floor position / box spring ~ 330mm) by simultaneously pressing the 2 keys variable height (continuous beep 5s).
6.6.4. Option I (Remote control Infrared) INFRARED ASSOCIATED WITH MOBILE CARER CONTROL (ACO) Back rest Variable height Electric folding legs Chair position 6.6.5. Option R (Mobile carer control ACO) To activate the function, press the « arrow » key (up or down) of the desired function.
6.6.6. Option N (Flexible arm control) Functions can be locked using the caregiver box. FLEXIBLE ARM CONTROL RELATED WITH MOBILE CARER CONTROL Electric folding legs Locking indicator Back rest light Variable height Chair position The flexible arm is fixed to one of the long sides of the bed observing a minimum distance of 63 cm from the panels or any accessory.
6.7. BRAKING 6.7.1. Separate braking Check that the wheels are locked by trying to move the bed. If this is not done, the patient or another person who leans on the bed may fall. 6.7.2. Centralized braking (option A) Central bed braking must be activated when the bed does not need to be moved. A : Braking: Press the pedal down with the foot.
6.7.3. Lateral head displacement (option T) To allow lateral movement of the bed, lift the bed in the high position. During the rise of the variable height, 1 stop occurs at a height of 680 mm, then an audible warning indicates that the bed on the head side is no longer braked.
7. OPERATION OF THE SLEEPING SURFACE 7.1. Back rest 7.1.1. Standard back rest In the event of a power cut or failure, flatten the back rest as follows : a) Disconnect the power supply. b) Dismantle the headboard . c) Stand at the head of the bed and take hold of the back rest handle with one hand. Push or lift to compensate the pressure exerted by the patient and unhook the clip ...
7.1.3. Emergency release of the back rest (Cardio Pulmonary Resuscitation) 1) Grasp the back rest with one hand. 2) With the other hand, activate one of the two handles on the back rest while lowering. If the handle is released, the back rest will stop moving. Release handles To reactivate the cardiopulmonary resuscitation system of high-load beds (270Kg), actuate the backrest electric actuator to its minimum position by pressing the...
➢ Removing a bed board : Assemble the highest board on the head side. 8.2. Metal side rails ➢ To positionning the side rail : Feet side Head side Insert amount of the side rail folded in the grip jaw on the head side of the bed to hear “CLIC”. ...
➢ To lower the side rail : Feet side Head side ➢ To raise the side rail : Check that the side rail is locked by trying to fold it without using the release pin 8.3. Wooden barriers See the attached instructions for fitting the wooden side rail. ➢...
8.4. Aluminum half side rails See the attached instructions for fitting the aluminum half side rails. ➢ To raise the side rail. 1-In intermediate position : Raise the top side rail with both hands until it locks. Check that it is properly slot in by trying to get it down without using the unlocking buttons. 2-In high position : ...
➢ To lower the side rail. Raise the top side rail. Press on the two unlocking buttons with both hands. Support the rail as it lowers. 8.5. Angled lifting pole and IV stand In case of psycho-behavioral problems of the patient, it is necessary to fix the stem to the bed with the screw and the washer provided.
9. SPECIFIC FUNCTIONALITIES. 9.1. Wall stop A663-00 9.2. Use of night light (option J) Possibility with all remote controls to use the night light fastened under the bed base. The function is switched on and off by simply pressing the back-rest function.
9.3. Battery (option K) The (optional) back-up battery enables you to perform 9 complete cycles when the bed is not connected to the mains. After these 9 cycles, a beep warns you that you need to to re-connect the bed to the mains to recharge the battery.
To resume the movement, release the button and press again or staying press 3s. WINNCARE recommends leaving a space of at least 50cm between the Trendelenburg headboard and the wall in the case of use of the Trendelenburg position with a bracket or IV pole.
9.5. Divisibility of bed base The bed AERYS-315 is divided into parts : Half bed base foot 57kg / Half bed base head 44kg The kit of transport A653-00 can divide the bed effortlessly. This accessory consists of 2 identical sets foldable. They are installed on trolley(1)”...
10. MAINTENANCE No maintenance should be done in the presence of a patient on the bed. 10.1. Identification Warning Manufacturer's address and Complies with regulations (EU) contact details Product reference and name Read the instruction manual Product serial number YY : Year of manufacture PP : Product empty weight MM : Month of manufacture XXXXX :Internal numbering...
10.3. Maintenance Quality control of medical beds will be made by technical staff or trained biomedical and taking into account the normal conditions of use specified in the user guide, on a bed equipped with its accessories, including its barriers if necessary. The bed must be available to perform all quality control at least once a year, but also on special request and corrective maintenance on the performance that could be affected by the intervention.
10.4. Quality inspection of medical beds ANNUAL MAINTENANCE OF MEDICAL BEDS IDENTIFICATION ESTABLISHMENT Type model / Trademark : Date of manufacture : Serial or park number : Application Environment Congruent Improper VISUAL CONTROL applicable Identification - label General cleanliness Condition and attachment of head and footboards Lifting pole (fastening, strap, winder, …) Sleeping surface (bedspring) Verification of tightenings, diverse nuts and bolts, pins, pivot...
10.5. Cleaning and disinfection High pressure cleaning, machine cleaning, washing tunnel or jet cleaning is prohibited. Unplug the mains lead. Check that all the electrical parts are connected together. All the sockets of the supply box must be used or filled, otherwise its watertightness is not guaranteed.
Do not rinse or wipe. Check the operation of all the bed functions Repair if necessary Pack in thermoplastic film Warning In the event of additional precautions (Contact precautions, Droplets or Air), apply the measures recommended by the hospital hygiene department The use of a Javel water solution of more than 5000ppm (0.5% of active chlorine) should be justified by a microbiological risk and only applied for the required time (Risk of ageing of some materials, especially their colour).
10.8. Troubleshooting Guide DEFAULT POSSIBLE CASE HELP The whole bed is not working The bed is not electrically powered Check the electrical connection (Green power LED on controlbox) The remote control is fully locked Unlock the remote control The infrared remote control is not powered or is not Remove the rear cover connected to the receiver.
Measures should therefore be taken to ensure that the bed is no longer used for the purpose it was originally intended. The crossed-out bin symbol indicates that the electrical elements of the Winncare product are considered to be Electrical and Electronic Equipment (EEE), subject to EU Directive 2012/19 / EU on the Selective sorting of electrical and electronic waste (DEEE);...