The Counters are incremented any time the calibration section or the Factory
parameters section have been modified.
At power on, the display will show the current software revision number followed
by the message of the Calibration Count "[AL[nt" then a number i.e. "123". The
number from the counter memory.
"PAr[nt" and probably a different number, i.e. "234".
reset to 0, they will increment until the display can no longer hold the values. (1 to
999999). It is expected we will never have more than 1 million calibrations in the
life of the machine.
Each display is held for 1-2 seconds.
The scale will then continue to do the display test and go to normal weighing.
If during the time the counting displays are shown, the user presses the [Tare] key,
the user will be given a message to enter the passcode necessary to calibrate the
scale, "P - - - - " Enter the code "P0000" to Enter calibration or "P1000" to enter
the parameters, followed by pressing the [Tare] key.
The Calibration access will allow user calibration (See section 15.1) and the
parameter code will allow access to the following parameters. (see section 15.2).
"F4 Int"
"F5 rEZ"
"F6 SCS"
"F7 Cnt"
"F8 Zem"
"F9 Lvd"
Then the Parameter Counter message of
Initial Zero Range
Re-Zero range
Successive Tare Enable
Display ADC counts
Zero Mode
Low voltage detection