Philips PM2528 Operating Manual page 50

Automatic rms multimeter
Hide thumbs Also See for PM2528:
Table of Contents

Advertisement

50
Resistance measurements (Q2w, 24w)
Select the INT, EXT or MAN start mode.
— Select the function Q2W or 24w.
= Select the correct range or the AUTO ranging mode.
Select the
HIGH RES mode or normal resolution mode.
— Connect the resistance to be measured to the correct input sockets.
Qaw
—0
(positive) and V-Q (negative)
Naw
— PROBE
Remarks
@ When only dashes are displayed (e.g. -_——.~— k Q) in the EXT or MAN start mode press START INT or
MAN,
@ Inthe Q4W function the 0 and the GUARD must be interconnected,
in the 24W function the 4-wire test lead can be used, For connector data, see fig. 15.
@ When very high resistances must be measured with high accuracy the input current has to be adjusted.
Refer to section 6.4.2. on page 45.
PROBE INPUT
MULTIMETER
PROBE PLUG
FRONT VIEW
FRONT VIEW
Fig. 15. Connection of the PROBE input for 24W measurements.
@ The relative reference mode can be used.
@ Use shielded cables in the high ranges to prevent errors in the measurements due to external influences.
@
In the 200M Q and 2000M &2 range the response time increases proportional to the measured value and
can be 22 seconds,
© Overload is indicated with OL in the display.
®@ In the table below, the ranges, ranging limits, scale lengths and measuring currents are represented,
Naw
©
<High resolution mode
Range
AUTO UP
AUTO DOWN
display
measuring
ranging
ranging
length
current
200.009
{2 | 220.000
240.006.
10mA
2000.08
.Q | 2200.00
0200.08:
2400.08:
1mA
20,0000k 2 | 22,0000:
02.0008:
24,0008:
100p A
200.000k Q | 220.000
020,000:
240.000:
102 A
2000.00k 2 | 2200.08
0200.00:
2400.08,
IBA
20,0000MS2 | 2200.08
02.0008
24,0008.
100n A
200.06M 2
220.00:
240.08:
0n A
2000.8M2.
2400.0
In A