Philips PM2528 Operating Manual page 17

Automatic rms multimeter
Hide thumbs Also See for PM2528:
Table of Contents

Advertisement

2.1.9,
V7
Excluding ranging:
synchronous start.
asynchronous start
Synchronous start means that the input signal and the start
command are given simultaneously,
For further information refer to the specification of the PM9211,
section 3.4.2. on page 26.
Peak voltage measurements (Vpeak, optional)
Ranges
Max, input voltage .
Measuring modes
Measuring functions
Measuring function
selection
Required optional type numbers
Measuring speed at internal start
Resolution
Accuracy at reference conditions
Temperature coefficient
Input impedance
Common Mode Rejection Ratio
(1k 9 unbalance and GUARD
connected to 0)
Maximum Common Mode Voltage
Frequency range
Maximum VHz product
NOTE:
600Vrms or 900V peak
Normal Mode (NM)
High Resolution Mode (HRM)
3% digit display
4% digit display
highest value (DC-coupled)
Vpeak minimum
:V
+
_, lowest value (DC-coupled)
Vpeak-to-peak'
:V-°
, peak-to-peak value (AC-coupled)
With switch at the front of the PM2528
1xpress=V
*
2xpress=Vi
v
3 x press = V
¢
Peak voltage plug-in unit PM9259
V "andV
vy
: 3.3 to 4 measurements/s
ve
: 1.6 to 2 measurements/s
100uV in the 2000mV range at
HRM.
1mV in the 2000mvV range at NM
+ (0.2% of reading + 0.2% of range)
10 Hz-
10kHz
+ (0.5% of reading + 0.2% of range)
10kHz - 100kHz
+ {1% of reading + 0.2% of range)
Vpeak maximum :V
*
Periodical signals:
Non-periodical signais:
+ (1% of reading + 0.2% of range)
See also section 6.6.1 on page 54.
+ (0.02% of reading + 0.01% of range} /° C
1MQ//55pF
120 dB for DC signals
80 dB for AC signals up to 50Hz, decreasing with 20 dB per decade
60Vrms, 85Vpeak between 0 and GUARD
250Vims, 400Vpeak between GUARD and earth.
Vv *andV
v
: 0-100kHz (DC - coupled)
ve
:10Hz - 100kHz (AC - coupled)
2.107
The meaning of the START modes in the Vpeak functions differ from those in the other functions.
Refer to section 6.6.1. on page 54,