Accuracy at reference conditions ,
valid between 1% and 100% of range
fe
Function V~
~ Function V >
Temperature coefficient
Input impedance
Common Made Rejection Ratio
(CMRR, 1k Qunbalance and
GUARD connected to 0}
Maximum direct voltage on
V*< for no additional error
AC detector
Maximum crest factor
Response time
Max, slew rate of input signal
Maximum input voltage
Maximum VHz product
Maximum Common Mode
voltage
Frequency range
Accuracy
30Hz
~— 20kHz
20kHz — 100kHz
+ (0.2% of reading + 0.07% of range}
+ (0.4% of reading + 0.07% of range}
30Hz
20kHz
| + (0.1% of reading + 0.05% of range}
200V
20kHz — 100kHz
| + (0.3% of reading + 0.05% of range)
and
100kHz — 300kHz
| +{1
%of reading + 0.05% of range)
2000V
300kHz — 500kHz
| + {5 %of reading + 0,05% of range}
AC component: same accuracy as V~ . Additional for DC
component: + (0.05% of reading + 0.01% of range)
£ (0.01% of reading + 0.01% of range}/°C
1M QU/ 55pF
DC signals
2 120dB
AC signals of 50Hz:
80dB; decreasing with 20dB/decade
25 x range
True RMS converter
end of range
So OE.
reading
e.g. at end of range the maximum crest factor is 4.5
Warning on the display when the crest factor is exceeded: i=
in the least significant digit.
Excluding ranging:
asynchronous start
synchronous start
NM
08s
05s
HRM
O7s
07s
Including ranging:
asynchronous start
synchronous start
Synchronous start means that the input signal and the start command
are given simultaneously (e.g, in EXT START mode).
1000V/ys for voltages > 200V
600 Vrms, 900 Vpeak
2.107
60 Vrms or
85 Vpeak between 0 and GUARD
250 Vrms or 400 Vpeak between GUARD and earth.