42
6.1.
6.2.
6.2.1,
OPERATION
SWITCHING ON
The instrument is ready for use after earthing and connecting it to the mains. It is switched on by means of the
push-button switch "POWER". A warming-up time of approx. 30 minutes should be taken into account to
meet the specified accuracy,
When the PM2528 is brought from a cold into a warm environment condensation can cause incorrect readings.
NOTE:
Before connecting an input signal to the PM2528 select the correct function and range.
CONTROLS, CONNECTORS AND
DISPLAY
Front panel
Ranging
UP
Up-ranging: 1 x press UP gives next higher range.
DOWN
Down-ranging: 1 x press DOWN gives next lower range.
AUTO
©
Automatic ranging: ranges are selected automatically.
LED in push-button lights at AUTO RANGING.
Measurements are started internally.
Measurements can be started externally via BNC or interface.
Measurements can be started by pressing the push-button. MAN.
LED in selected START mode lights.
Function
O
Direct voltage measurements.
©
Alternating voitage measurements
O.
Alte:
i n g
voltage measurement:
9 direct voltage component.
©
Two-wire resistance measurements.
©
Four-wire resistance measurements, possible with 4-wire test lead
©
Direct current measurements.
©
Alternating current measurement including de component.
O
Peak voitage measurements. Only possible with peak voltage option
PMg259.
O
High frequency voltage measurements. Only possible with HF option
PM9258 and HF probe PM9211.
©
Temperature measurements. Possible with Pt-100 probe e.g. PM9249.
NOTE: ©
Function is switched on when LED in push-button lights.