30
3.6.
ANALOG OUTPUT PM8254
The analog output PM9254 offers the possibility to monitor a part of the displayed value. The output voltage
sweep corresponds with a selectable number of digits of the display, and is galvanically separatéd from the
input. The analog output PM9254 consists of two plug-in cards.
Output voltage
Accuracy
Temperature coefficient
Output sockets
Output protection
Output system
Output resistance
Maximum voltage between
input (0 terminal) and output
Response time
Output modes
Output mode selection
Resolution
O-1V
+ (0.2% of reading+ 0.1% of range)
excluding PM2528.
£ (0.025% of reading + 0.005% of range} per °C,
excluding PM2528.
HI and LO; LO connected to earth
Short circuit protected
Max. input voltage 15V.
Digital-to-Analog Converter.
2002
250Vrins, 350Vpeak
< 0.5 s, excluding PM2528
selected digits
D5 = most-significant digit
DO = least-significant digit
* scale length is 2400
via switch at the rear of the PM2528.
mode A
1:2500
other modes
range for
output voltage
ov
=
1V
0000. .— 2500...
-000..—.999..
-500..—.499..
«000
. 999,
-.500,—..499,
---000—...999
+ --500 ~...499