Time Specification; Mains Supply; Mains Supply Voltage; Mains Supply Frequency - Philips PM2528 Operating Manual

Automatic rms multimeter
Hide thumbs Also See for PM2528:
Table of Contents

Advertisement

18
2.2.
2,3,
2.3.1.
TIME SPECIFICATION
Measuring speed
Input settling time
Measuring time
Sampling time
Response time
Range response time
Recovery time from overload
MAINS SUPPLY
Mains supply conditions
Mains supply voltage
Reference value
Rated range of use
Normal Mode (NM);
3.3to
5.4 measurements/s
High Resolution Mode (HRM);
1.6 to
2.7 measurements/s
oo
High Speed Mode (HSM);
13
to18
measurements/s
'
(The measuring speed depends on the measured value)
Depending on selected function and measuring speed; automatically
f
selected.
i
Depending on selected function and measuring speed,
50Hz version
60 Hz version
1x83 1/3ms
2x 83 1/3ms
1 x 46 2/3ms
Depending on selected function.
Depending on selected function.
overload
:
to 200V
to 1000V
ia
i
acc. to JEC 359 — group 2.
220 V + 1%
:
220 V + 10%
nD
NOTE:
The instrument can be adapted to nominal supply voltages of 110V, 120V, 240V.
Refer to section 5.1.3. on page 41.
Mains supply frequency
Reference value
Rated range of use
NOTES:
50H2 + 1%
SOHz + 1%
The instrument can be adapted to a nominal mains frequency of 60Hz.
Refer to section 5.1.3. on page 41.
Mains supply interruptions
Interruptions
<30 ms : no influence
between 30ms and 500ms
: instrument may either restart or continue.
> 500ms: instrument wili restart. Conditions are identical to the
situation after switching on.