Philips PM2528 Operating Manual page 46

Automatic rms multimeter
Hide thumbs Also See for PM2528:
Table of Contents

Advertisement

46
6.4.4,
Example:
— Select FUNCTION
Q2W
— Select START INT
— Select range 2002 at manual ranging
— Select HIGH RES mode
{LED on)
— Switch the OFFSET mode off
(LED off}
— Connect a resistor R {see table below) between the O and V-Q socket
Display value
Remarks
100.000 $2
000.0002
press OFFSET, LED lights
050,0002
value 50 © below reference value
060.0002
value 60 Q above reference value
OL
range is 20092
000.0002.
100.0002
press OFFSET, LED off
High resolution mode (HIGH RES)
By pressing the push-button HIGH RES the PM2528 is switched into or out of the high resolution mode.
tn the HIGH -RES mode the display is extended by one digit. In the table below the display length is given,
Normal mode
digits
4%
HIGH RES mode
digits
Function
ranges up to and including 20 MQ.
ranges 200M {2 and 2000M S2
fous