4-10.
VERIFY
4-11.
VERIFY allows you to determine if the contents of either RAM bank
ls maintained for an extended period of time. This test should only be run
after RAMTEST. It compares the data stored in the PMS under test to the
pattern written by RAMTEST.
4-12.
POWER
4-13. POWER executes a worst case power drain loop in an PMS. It is
used in conjunction with troubleshooting procedure in making worst case
current measurments.
4-14,
ADRTEST
4-15.
ADRTEST is a continuous loop routine that writes a byte of data to
every location in both banks of RAM. This routine aids in troubleshooting
the RAM address circuitry by toggeling all of the address lines.
UL-16.
SELTEST
given data string to and from the PMS in a known repeating pattern. This
then generates the signatures used in signature analysis.