SECTION
III
Disassembly and Reassembly
3=5).
Opening the case (procedure 3-1).
Removing the Logic board (procedure 3-2).
Removing the Connector/Cable assembly (procedure 3-3).
Removing the Battery (procedure 3-4).
Assembling the Logic board and the Connector/Cable assembly (procedure
Assembling the case (procedure 3-6).
Oo
Oo
CAUTION
The ICs in the HP 82T713A are easily damaged by
ESD (Electro-Static Discharge) if adequate
precautions are not taken. Observe these
precautions:
The surface of the repair bench must be grounded.
Any tool used in the assembly, disassembly, or
troubleshooting of the PMS must be grounded
to the surface of the repair bench.
All parts must be kept within their
antistatic containers during storage and while
being carried to the repair bench.
While handling parts and assemblies, repair
personnel must wear a wrist strap grounded
to the surface of the repair bench.
Use an air ionizer to drain static charge from
the repair bench area.