Guidance and Manufacturer's Declaration – Electromagnetic Immunity
POC-615-12 SERIES is intended for use in the electromagnetic environment specified
below. The customer or the user of the model POC-615-12 SERIES should assure that it is
used in such an environment.
Immunity test
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Interruptions and
voltage varia-
tions on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the A.C. mains voltage prior to application of the test level.
POC-615-12 Series User Manual
IEC 60601 test
Compliance
level
level
±8 kV contact
±8 kV contact
±15 kV air
±15 kV air
±2 kV for
±2 kV for power
power supply
supply lines
lines
±1 kV for input/out-
±1 kV for input/
put lines
output lines
±1 kV line(s) to
±1 kV line(s) to
line(s)
line(s)
±2 kV line(s) to
±2 kV line(s) to
earth
earth
Voltage Dips:
Voltage Dips:
i) 0% reduction
i) 0% reduc-
for 0.5 cycle at
tion for 0.5
50Hz,
cycle at 50Hz,
0% reduction
for 1 cycle at 50Hz
tion for 1 cycle
Performance
at 50Hz
Criterion B
ii) 70% reduction
Performance
for 25/30 cycles at
Criterion B
50/60Hz
ii) 70%
Performance
reduction for
Criterion C
25/30 cycles at
50/60Hz
Voltage Interrup-
tions:
Performance
0% reduction
Criterion C
for 250/300 cycles
at 50/60Hz
Voltage Inter-
Performance
ruptions:
Criterion C
30A/m
tion for 250/
300 cycles at
50/60Hz
Performance
Criterion C
30A/m
18
Electromagnetic environ-
ment –guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Main power quality should be
that of a typical commercial or
hospital environment.
Main power quality should be
that of a typical commercial or
hospital environment.
Main power quality should be
that of a typical commercial or
hospital environment. If the
user of the model POC-615-
12 requires continued opera-
tion during main power inter-
ruption, it is recommended
that the model POC-615-12
0% reduc-
be powered from an uninter-
ruptible power supply.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commer-
cial or hospital environment.
0% reduc-