Guidance and manufacturer's declaration - Electromagnetic
immunity
The arm blood pressure monitor is intended for use in the electro-
magnetic environment specified below. The customer or the user of
arm blood pressure monitor should assure that it is used in such an
environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power frequency
Magnetic field
IEC 61000-4-8
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
NOTE: UT is the a.c. mains voltage prior to application of the test level
IEC 60601
test level
±8kV contact
±2 kV, ±4kV, ±8 kV,
±15kV air
Not applicable
Not applicable
Not applicable
30A/m, 50/60Hz
Not applicable
10 V/m
80 MHz - 2,7 GHz
80 % AM at 1 kHz
Compliance level
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Not applicable
Not applicable
Not applicable
30A/m, 50/60Hz
Not applicable
10 V/m
80 MHz - 2,7 GHz
80 % AM at 1 kHz
English
45