Learning The Controls - Tektronix 2205 Operator's Manual

Hide thumbs Also See for 2205:
Table of Contents

Advertisement

Operation
LEARNING THE CONTROLS
If you have not read Section 1 yet, you should do so now. Then,
after turning the power on, let the oscilloscope warm up for a
few minutes before starting this procedure.
1. Set instrument controls as follows:
Display
INTENSITY
FOCUS
Vertical (both channels)
POSITION
MODE
VOLTS/DIV
VOLTS/DIV Varible
Input Coupling
Midrange
Set for clearly defined trace
Midrange
CH 1
.1 V (iX)
CAL detent (fully clockwise)
AC
Horizontal
POSITION
MODE
SEC/DIV
SEC/DiV Variable
Midrange
X1
0.2 ms
CAL detent (fully clockwise)
Trigger
SLOPE
LEVEL
MODE
SOURCE
2. Connect a signal adapter to the Channel 1 input BNC con¬
nector (labeled CH 1 or X). Attach the ground lead to the
collar of the EXT INPUT connector and apply the signal
adapter tip to the PROBE ADJUST terminal.
If necessary,
adjust the TRIGGER LEVEL control to get a stable display.
3. Change the Channel 1 input coupling to GND and use the
Channel 1 POSITION control to align the baseline trace to the
Midrange
P~P AUTO
CH 1
2-6
2205 Operator's

Advertisement