Ametek MAGNETROL B40 Instruction Manual And Replacement Parts

Ametek MAGNETROL B40 Instruction Manual And Replacement Parts

High pressure –high temperature liquid float level switch

Advertisement

Quick Links

B40
High pressure –high temperature liquid float level switch
inSTRUcTiOn ManUaL anD REpLacEMEnT paRTS
DEScRipTiOn
Magnetrol's B40 liquid level switch is specifically designed
and constructed for high pressure, high temperature service
conditions. (It is suitable for use on 0.65 minimum specific
gravity liquids).
OpERaTing pRincipLE
B40 level switches employ permanent magnetic force as the
only link between the float and the switching element. As
the pivoted float follows liquid level changes, it moves a
magnetic sleeve ¨ into or out of the field of a switch
actuating magnet ≠ causing switch operation.
A non-magnetic barrier tube Æ effectively isolates the switch
mechanism from the controlled liquid.
Mercury switch
z
x
pivots
Unpacking
Unpack the instrument carefully. Inspect all units for dam-
age. Report any concealed damage to carrier within 24
hours. Check the contents of the packing slip and purchase
order. Check and record the serial number for future refer-
ence when ordering parts.
pivot
Tension spring
y
Swing-in position
Float
agEncY appROVaLS
Agency
Approval
II 2G EEx d II C T6, explosion proof
ATEX
CCE ¿
R1 (1) 136/MI/433, explosion proof
Class I, Div. 1, Groups C & D
FM
Class II, Div. 1, Groups E, F & G, Type NEMA 7/9
FM/CSA ¡
Non-Hazardous area
Explosion proof area –
Groups B, C, D, E, F & G Type NEMA 4X/7/9
Explosion proof area
SAA ¡
LRS
Lloyds Register of Shipment (marine applications)
GOST/
Russian Authorisation Standards
GOSGORTECHNADZOR ¡
Other approvals are available, consult factory for more details
¿ For CCE approved units, use the ATEX explosion proof model
numbers.
¡ Consult factory for proper model numbers.

Advertisement

Table of Contents
loading

Summary of Contents for Ametek MAGNETROL B40

  • Page 1 High pressure –high temperature liquid float level switch inSTRUcTiOn ManUaL anD REpLacEMEnT paRTS DEScRipTiOn Magnetrol’s B40 liquid level switch is specifically designed and constructed for high pressure, high temperature service conditions. (It is suitable for use on 0.65 minimum specific gravity liquids).
  • Page 2: Model Identification

    MODEL iDEnTiFicaTiOn a complete measuring system consists of: 1. Code for high temperature / high pressure liquid float level switch in an external cage BASIC MODEL NUMBER B 4 0 cage mounted float switch for high temperature / high pressure applications Down to S.G.
  • Page 3: Installation

    inSTaLLaTiOn piping MOUnTing Figure 3 shows a typical piping installation of a Model B40 Before welding, adjust piping to bring control to a vertical in a pipeline to a pressure vessel. Reference lines, on float position. B40 level controls must be mounted within three chamber, should be aligned to correspond with liquid level (3°) degrees of vertical.
  • Page 4 WiRing nOTE: If control has been furnished with an explosion proof Most mechanical control switch housings are designed to allow 360° positioning of the cable entries by loosening the (cast) or moisture proof (gasketed) switch housing, check set screw. See figure 4. On high temperature applications the following: (above 120°...
  • Page 5: Preventive Maintenance

    pREVEnTiVE MainTEnancE periodic inspections are a necessary means to keep your Magnetrol level control in good working order. This control is, in reality, a safety device to protect the valuable equipment it serves. Therefore, a systematic program of "preven- tive maintenance" should be implemented when control is placed into service. if the following sections on "what to do"...
  • Page 6: Replacement Parts

    REpLacEMEnT paRTS Model part number code item Description B40-1F30 B40-1B60 B40-2F30 B40-2B60 B40-5c20 B40-3c30 B40-4c40 Housing Cover Refer to Bulletin 42-783 or 42-794 Switch Mechanism Jam Nuts(s) 10-2107-001 (Qty. 4) Attraction Sleeve 32-4203-001 32-4902-001 32-4203-001 04-5411-002 E-Tube Gasket 89-5955-001 89-5920-001 89-5951-001 Enclosing Tube 89-5920-001 E-Tube Kit...
  • Page 7 DiMEnSiOnaL SpEciFicaTiOnS in mm (inches) ø X Switch Types inches inches inches inches M20 x 1,5 or 1" NPT Weatherproof- ATEX - FM (NEMA 7/9) - F or S 10.12 1.66 5.93 4.29 (2 entries - 1 plugged) (Cast Aluminium) not for FM (NEMA 7/9) Weatherproof- 3/4"...
  • Page 8 All shipments returned to the factory must be by prepaid transportation. Magnetrol will not accept collect shipments. All replacements will be shipped FOB factory. BULLETIN N°: BE 46-602.7 EFFECTIVE: JANUARY 2005 UNDER RESERVE OF MODIFICATIONS SUPERSEDES: March 1997 European Headquarters & Manufacturing Facility Heikensstraat 6 9240 Zele, Belgium Tel: +32-(0)52-45.11.11 e-mail: info.magnetrolbe@ametek.com www.magnetrol.com...

Table of Contents