7

Technical specifications

Measurement mode
Measuring range
Measuring accuracy
Range 0 to 1 ‰
Range > 1 to 2 ‰
Range > 2 ‰
Range 0 to 1 ‰
Range ≥ 1 ‰
Ambient conditions
Operating temperature range
Storage temperature range
Storage temperature (sensor)
Ambient pressure
Ambient humidity
Minimum exhalation volume
Batteries or accumulator (optional)
Charging station (optional)
Charging time
Measurements per battery
pack/accumulator charging
Device interface
Doc no.: 059-07-1001779_A6020plus_WL_GB-5.2.1 / 12.2021
:
active and passive
:
0 to 5.5 ‰ BAC*
:
Max. error tolerance in temperature range from -
5°C to 45°C
:
± 0.05 ‰ absolute
:
± 5% of measurement value
:
± 10% of measurement value
Reproducibility with ethanol
standard in temperature range from -5°C to
45°C
0,017 ‰
1,7 % of measurement volume, the higher
volume is valid
:
-10°C to +50°C
:
-25°C to +70°C
:
0°C to +25°C, optimal
:
600 to 1400 hPa
:
20 to 98% rel. humidity
:
1.2 litres
:
4x LR6 (AA) alkaline Mignon batteries
(Industrial by Duracell recommended) or
Honeywell Healthcare Solutions GmbH -Akku
NiMH 4.8V (4 x 2100 mAh)
:
protection class II
Input: AC 100–240 V/50–60 Hz/93 mA
Output: DC 5.6 V/500 mA/2.8 VA
:
approx. 4 hours
:
approx. 2000 measurements, depending on the
batteries and the ambient conditions
:
USB and RS-232 interface
Data can be transmitted to a PC.
Part no. 1002249 and 1002251: Wireless
interface (IEEE 802.15.1) for usage with
wireless printer (part no. 1002255).