Designated use and description

The AlcoQuant 6020 plus is designed for a quick and accurate determination to check the
presence of alcohol or alcohol intake and is the optimal solution for the following applications:
police routine alcohol verification
testing of prison inmates
ensuring safety at work
The AlcoQuant 6020 plus determines the breathed air volume while expiration procedure and
takes a sample after 1.2 litres (standard setting).
The AlcoQuant 6020 plus contains a special electrochemical sensor, which reacts only to
alcohol. The unique characteristic of this sensor is its specific reaction to alcohol as well as
its
long-term
6-month calibration interval.
The influence of other substances that may be contained in the expiration air, such as
acetone, eucalyptol, etc. is negligible owing to the electrochemical measuring principle
applied. The sample collection system, which has been developed for maximum precision,
permits short measuring intervals, thus facilitating extremely short waiting periods between
measurements.
Thanks to the use of the new, improved ethanol sensor E 100 and the innovative sample
collection system, a high degree of reproducibility of measuring results is guaranteed. State-
of-the-art technologies for the prevention of faulty measurements and to ensure maximum
measuring precision are implemented in the device. To achieve this, the following functions
and components are constantly controlled.
Device temperature
Sample collection
Signal generator
LEDs
The device comes with a memory module and a universal interface (USB + RS-232). This
ensures that the device is completely backwards compatible. The data can be transferred to
a computer with the >Datamanager Alco 60XX< software and the data cable and analysed.
The device is also available with an optional wireless interface which can be used exclusively
to connect to an approved wireless printer (article no. E1002255).