Error messages and removal of errors

Each device error is signalised acoustically by a short dual signal and optically either by the
yellow
(error that does not impair measurement) or by the red LED (error impairing measurement). It
leads to a corresponding error message on the display.
This error message must be confirmed by briefly pressing >OK<.
Phase
Display on
device
Operation
ERROR 000 -
ERROR 100
Operation
Error C001 -
C100
Switching on Error
Calibration
Switching on Error
Buzzer
Switching on Error
Temperature
Switching on Error
LED
Switching on Error
Hardware
Switching on Calibration
expired
Doc no.: 059-07-1001779_A6020plus_WL_GB-5.2.1 / 12.2021
Result
LED signals Cause/solution
Device
RED
switches
off after
confir-
mation
Error
RED
sound
Return to
menu
Confirm
RED, then
with key
blinks
permanently
Confirm
YELLOW
with key
Confirm
YELLOW
with key
Confirm
YELLOW
with key
Confirm
RED,
with key
remains on
permanently
Confirm
YELLOW
with key
AlcoQuant 6020 plus
Replace batteries; if problem
persists the device is faulty.
Contact service.
Error in program sequence
Update required.
Contact service.
Device must be recalibrated.
No measurements are possible.
Contact service.
Buzzer defective.
Measurements are still possible.
Contact service, if necessary.
Device temperature is outside of
the permitted temperature range.
Warm or cool the device
accordingly.
The red, green and/or yellow LED
is faulty.
Measurements are still possible.
Contact service.
Replace batteries; if problem
persists the device is faulty.
No measurements are possible.
Contact service.
Calibration period expired.
Device must be recalibrated.
Measurements still possible, but
level of accuracy cannot be
guaranteed.
Contact service.