Maintenance

Calibration
The device must be calibrated every 6 months. This involves the following points:
Calibration of the device
Test whether device functions are in proper working order
Test of calibration
The device may only be serviced by Honeywell Healthcare Solutions GmbH or
service personnel who have been trained by Honeywell Healthcare Solutions
GmbH.
Please contact your dealer or send the device and all of its accessories to the following
address for servicing:
Honeywell Healthcare Solutions GmbH
Service
Alter Holzhafen 18
23966 Wismar, Germany
envitec-service@honeywell.com
To speed up the repair process, please obtain an RMA (Return Material Authorisation)
number. You can acquire an RMA from our website: http:\\envitec.com\.
Please state your RMA number in all your correspondence!
Cleaning
Please use only an impregnated disposable cloth with mild cleaning agents to clean the
device without excessive pressure
Disinfection
Please use only an impregnated disposable cloth with Isopropyl-Alcohol to clean the device
without excessive pressure (EN 60601-1) (Check of the permanency of the inscription).
Keep battery cover closed while cleaning and disinfection. Humidity must not enter
the holes of the device.
The electrical contacts inside the device (battery/accumulator) and at the USB port
must be clean and dry at all times to ensure that the device will function properly.
If alcohol has been used for cleaning/disinfection, a waiting time of min. 15 minutes
until the next measurement must be maintained!
Doc no.: 059-07-1001779_A6020plus_WL_GB-5.2.1 / 12.2021