3.2

Switch-on / Switch-off

Once the device has been switched on, it will automatically test all of its internal functions
and components. While performing these tests, the screen displays information about the
"next calibration" and "software revision status" for about 2 seconds.
Hold down the button for around 1 second.
Within 10 seconds the device is now ready to start measuring
(green LED blinking at an interval of 1 second).
Example: Information displayed while the unit is being switched on:
Switch-off
 Press key for approx. 3 seconds
If no measurement or key entry is made within the time set in the "Auto off" menu
item, the AlcoQuant 6020 plus turns itself off automatically.
Doc no.: 059-07-1001779_A6020plus_WL_GB-5.2.1 / 12.2021
The device is at the >Active measurement mode< menu item when
switched on.
The device is ready for use. The green LED blinks every second to
indicate readiness for start of the measurement.
The "Active/Passive" measurement modes can be switched
between in the operating menu or with the right softkey.
AlcoQuant 6020 plus
9