11.2.2 Entering the known tare in the numerical form
< Ptare > < manuAl >
The entered tare value will be used until you enter a new one. To delete it,
enter a zero value or confirm the menu item <clear>, pressing ➔.
11.3 Data-hold function
TPCJ-BA-e-2310
Display the <Ptare> menu setting and confirm, pressing
➔.
Use the navigation buttons to choose the setting <ma-
nuel> and confirm by ➔-button.
Enter the known tare value; for introducing the numerical
value see chapter 0, the active item is blinking.
The entered weight will be saved as tare; <NET> and
<PTARE> symbols will be shown and tare weight with a
negative symbol will be displayed.
Place a filled weighing container.
Wait until the stabilization indicator is displayed (
Read out the net weight.
Choose the menu item <hold>.
Place the weighed material.
Confirm by ➔-button.
The first stable weight value is held symbolized by [HOLD]
in the upper edge of the display. After unloading, the value
is held in the display for another 10 s.
).
35