PICARRO G2302 User Manual page 38

Table of Contents

Advertisement

1. Using the 1.5mm hex driver, remove analyzer top lid by removing six (3
per side) M3 x 6mm socket flathead screws.
2. Remove the piece of foam from around the input bulkhead by sliding it
towards the back of the analyzer.
3. Using the 5/8" wrench, loosen the retaining nut on the input bulkhead
(about 1 full turn should be enough).
Foam
covering
filter
4. Slide the filter cover (with foam on top and side) towards the right side
of the analyzer to remove it.
5. Using the 9/16" and 11/16" wrenches, disconnect the filter from the
tube section near the front of the analzyer.
Picarro G2302 Analyzer User's Guide
Rev. B 1/14/11
Remove
foam
View of right side of DAS
Bulkhead Retaining Nut loosened
Retaining
Nut
38