INSTALLATION

This section describes the setup and installation of the Picarro Analyzer. Please
read and understand this section thoroughly before proceeding with the
installation.
WARNING: Do not attach electrical power to, or start the analyzer
until after attaching and turning on the External Vacuum Pump. Do
not disconnect the vacuum line while the analyzer is running.
Failure to do so could result in damage to the optics.
WARNING: Picarro sells USB enabled devices, such as GPS, which
are approved for use. Please do not connect USB hubs or
unapproved USB devices, other than flash drives to the computer
because they can interfere with the operation of the analyzer.
WARNING: If rack mounted, the Analyzer cannot support itself using
a front rack mount kit alone. The instrument must be supported by a
shelf or additional rails attached to the rack.
CAUTION: It is imperative that the analyzer have adequate
ventilation and/or cooling to maintain the ambient temperature below
35 ° C when operating. Failure to provide adequate a irflow, especially
clearance at the front and rear panels, to ensure proper airflow
and/or cooling to the analyzer will result in overheating of the
analyzer causing a shutdown and potential damage. There should
be 4" (10cm) of clearance in the front and back of the analyzer.
Thermal
Specifications
Ambient Operating
Temperature
CAUTION: If the analyzer has been stored at less than 10 °C, allow
the components to equalize to room temperature before starting the
installation process.
NOTE: Take care to ensure that warm air is exhausted from any
enclosure in which the analyzer is mounted.
Picarro G2302 Analyzer User's Guide
Rev. B 1/14/11
Min
Max
Description
10 °C
35 °C
Worst-case environmental limits
(unless otherwise specified)