Basic Analyzer Setup

1
Remove the Analyzer and the External Vacuum Pump from their respective
shipping containers.
2
Place the Analyzer on a bench top or flat surface. Place the External
Vacuum Pump near-by or on the floor. Don't push the analyzer into position
yet, there are cables to be installed at the back.
3
Unpack the analyzer accessories. (gas line, AC power cables, manual,
certificate of compliance, and USB Flash Drive). The Certificate and USB
drive should be stored in a safe place. They may be required if you contact
us with questions about your analyzer.
4
Remove the caps from the Analyzer gas connection inlet and vacuum
connection ports. Save the caps for later use, they should be reinstalled if
the analyzer is stored, moved or shipped.
5
Remove the caps from External Vacuum pump. Save the caps for later use.
They should be reinstalled if the pump is stored, moved or shipped.
6
Attach the gas line between the Analyzer vacuum port and the External
Vacuum Pump. See Figure 1. Hand tighten the nut, then make an
th
additional 1/4
7
Connect the AC power cable to the External Vacuum Pump but do not plug
it in.
8
If desired, attach a tube to the External Vacuum Pump exhaust port and
direct to a safe place for venting the mixture of sample gases.
Picarro G2302 Analyzer User's Guide
Rev. B 1/14/11
turn with an 11/16" wrench (not supplied).