Calibration point #1
Calibration point #2
Calibration point #3
Calibration values are input into the software by selecting the "User
Calibration" from the Tools menu and entering the slope and intercept for
each species. This is a password-protected function, with the default
password "picarro" and can be changed by modifying an *.ini file as
previously described. After the calibration is entered, it will take effect
immediately after clicking "ok."
To return to the factory calibration, simply set the slope to 1 and the
intercept to 0 for each species.
Picarro G2302 Analyzer User's Guide
Rev. B 1/14/11
Value given
Value of calibration
by analyzer
standard
200.1
202.7
600.3
597.6
400.0
400.0