7.3.3
"Power Manager" main menu
Function
DMPM ...
7.3.4
"Other Options" main menu
Function
ICS Fan Operation
OSD Settings
RadiForce LX300W
Instruction for Use, 03/2014
Adjustment / setting
range
DMPM Lamp Dimmed
DMPM Lamp Off
DMPM Disabled
Default: DMPM Lamp
Dimmed
Adjustment / setting
range
Standard
Operating Room
Default: Standard
Horizontal 0 ... 139
Default: 75
Vertical 0 ... 41
Default: 16
Transparency 0 ... 255
Default: 255
7.3 Description of OSD menu
Description
Setting the DMPM mode
•
Lamp Dimmed: Backlight brightness is re‐
duced to a minimum. This saves energy
and the lamp only requires a short warm-
up time after being reactivated.
In this mode, the energy saving mode
time counts toward the operating hours.
•
Lamp Off: The backlight is shut down
completely. The energy saving mode is
increased compared to "Backlight Dim‐
med".
Prior to diagnosing radiographic images
however, the lamp requires a warm-up
time after reactivation.
•
Disabled: DMPM signals are ignored. The
monitor does not change to energy saving
mode.
Description
Fan control
•
In standard mode, the fans run at half
speed. Above a specific ambient tempera‐
ture, the fans run at full speed.
•
The fans are switched off in Operating
Room mode. Above a specific ambient
temperature, brightness is reduced by
half.
Setting the position and transparency of
the OSD menu
Horizontal and vertical coordinates establish
the position of the OSD menu.
Note: The OSD menu cannot overlap a PiP
window or an image from the Fail-Safe func‐
tion. Corresponding entries are suppressed
or the OSD menu position is corrected auto‐
matically.
Use "Transparency" to change the transpar‐
ency of the OSD background.
Operation