Operation
7.3 Description of OSD menu
Function
Position / Size (PiP)

7.3.3 Power Manager menu

Function
DMPM ...
32
Adjustment / setting
range
H-Position
Default: 64
V-Position
Default: 64
H-Size
Default: 96
V-Size
Default: 80
Adjustment / setting
range
DMPM Lamp Dimmed
DMPM Lamp Off
DMPM Disabled
Default: DMPM Lamp
Dimmed
Description
Establishing the PiP display
You set the position and size of the picture-
in-picture display here. The zero position is
the upper left corner of the window.
Description
Setting the DMPM mode
Lamp Dimmed: Backlight brightness is re-
duced to a minimum. This saves energy
and the lamp only requires a short warm-
up time after being reactivated.
In this mode, the energy saving mode
time counts toward the operating hours.
Lamp Off: The backlight is shut down
completely. The energy saving mode is
increased compared to "Backlight
Dimmed".
Prior to diagnosing radiographic images
however, the lamp requires a warm-up
time after reactivation.
Disabled: DMPM signals are ignored. The
monitor does not change to energy saving
mode.