July 2023
Figure 2-4: Pneumatic installation, Rosemount OCX8800 with
reference air set, solenoids, and autocalibration, without COe
zero function
A. Sensor housing
B. Eductor air in
C. Electronics housing
D. Calibration gas in
E. Reference air in
F. Dilution air in
G. Dilution air flow meter 0.1 scfh
H. Instrument air out
I. Calibration gas flow meter (7 scfh, 20 to 30 psig [1.4 to 2.1 barg]
recommended)
J. Two-stage regulators
K. Instrument air supply
L. Pressure regulator/filter 35 psig (2.4 barg) for general purpose, 45
psig (3.1 barg) for hazardous areas
M. 2-in. pressure gauge 0-60 psig (0 to 4.1 barg)
N. Combination filter-reg. 0-60 psig (0 to 4.1 barg)
O. Flow meter 1-10 scfh
P. Flow meter 0.05-0.5 scfh
Q. Calibration gas out
Quick Start Guide
Quick Start Guide
17