Menu; Navigation In The Menu - KERN MCC-NM Service Manual

Medical chair scale
Table of Contents

Advertisement

5 Menu

Access to service menu „tCH"is locked in verified balances.
To disable the access lock, destroy the seal and actuate the
adjustment switch.For position of adjustment switch, see chap. 15.
Attention:
After destruction of the seal the weighing system must be re-verified
by an authorised agency and a new verification wire/seal mark fitted
before it can be reused for applications subject to verification.

5.1 Navigation in the menu

Call up menu
Select function
Change settings
Exit menu/
Return to weighing
mode
MCC-NM-SH-e-1810
In weighing mode, press
[F1 oFF] will be displayed.
With help of
selected one after the other.
Confirm selected function by
will be displayed.
Select desired setting by
the balance returns to the menu.
Press
and the scales will return to weighing mode.
and the first function
, the individual functions can be
.The current setting
and confirm with
,
15