After successful repair the balance will have to be reverified before it can be used again in a statutorily regulated field. 2 Introdution This service manual covers the MWB series and is edited for the authorized servicing personnel. Note all rights are reserved. Copying any part of this manual is prohibited without our permission.
3.2 Overview display Display Designation Description GROSS Gross weight display Lights up during indication of the gross weight Net weight display Lights up during indication of the net weight Illuminated after weighing scale was tared ZERO Zeroing display Should the balance not display exactly zero despite empty scale pan, press the button.
3.3 Overview Keyboard Button Designation Function ON/OFF button Turn on/off HOLD button Hold function BMI key Calculation of the Body Mass Index In menu: Confirm selection For numeric entry: Confirm numerical value In menu: Function key Call up menu ...
Information is available on KERN’s home page (www.kern-sohn.com) with regard to the monitoring of balance test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.
5.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances. 6 Transport and storage 6.1 Testing upon acceptance When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.
7 Unpacking, Setup and Commissioning 7.1 Place of installation / place of operation The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
8 Menu 8.1 Navigation in the menu Call up menu In weighing mode, press and the first function [F1 oFF] will be displayed. Select function With help of , the individual functions can be selected one after the other. Change settings ...
Page 13
Gravity Gravity not documented MWB-SH-e-1810.docx...
9 Error Messages Display Description Zero range exceeded (on start-up or when pressing the key) Load on weighing pan Excess load, during zero setting of weighing scale Incorrect adjusting process Fault on load cell Value outside the A/D converter range ...
10.3 Servicing, maintenance The appliance may only be opened by trained service technicians who are authorized by KERN. Disconnect the scales before opening. 10.4 Disposal Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
Page 16
The weighing result is The display of the balance is not at zero. obviously incorrect Adjustment is no longer correct. Great fluctuations in temperature. The balance is on an uneven surface. Electromagnetic fields / static charging (choose different location/switch off interfering device if possible) Should other error messages occur, switch balance off and then on again.
1. Carry out adjustment as closely as possible to admissible maximum load of weighing scale. Info about test weights can be found on the Internet at: http://www.kern-sohn.com. Observe stable environmental conditions. For warm-up time required for...
MAINTENENCE 13.1 General If the scale does not operate properly, find out the problem as possible. Determine whether the problem is constant or alternate. Be aware that problems can be caused by mechanical or electrical influences. Check the following. Water ...
13.5 Load Cell Connections Signal Green Supply + Signal Black Supply - White Shield Measuring Points Resistance Red (+ Exc) to White ( –Exc) 420 ±20Ω Green (+Sig) to Black ( –Sig) 350Ω ±5Ω Load cell connection for the 4 load cell’s.
13.6 Check PCB Voltages If the problem is in the PCB, use a multimeter to check the following voltages 13.7 PCB Input Voltage Check the PCB input power connecter voltage Voltage must be minimum 9VDC in to the pin AD+ 13.8 Check Battery Voltage Check the Battery Voltage The battery voltage below the 1.5VDC, replace the battery and install new...
13.9 Replace Main Board Release 4 pcs screw from bottom of the scale by using cross screw driver. Open the top cover, the main board lie on top cover, use cross screw driver loose 4 screw for main board, pull out all connector on the main board, then you can bring out main board carefully.