6.KEY NOTES FOR USER

PRIOR TO THE VERY FIRST USE
• Every operator is advised to be thoroughly familiarized with the content of this user manual
before starting up the device to ensure correct use of this device.
• Labomed trained installation person or trained authorized representative of Labomed dealer
will install the system. Please make sure that the following requirements continue to be met
for further operation:
• The connecting components have been properly connected. The screw connections have
been firmly tightened.
• The device is connected with the power cord supplied. The power cord being used is the
one designed for use with this system.
• When using drapes for the surgical microscope, make sure to remove the protective cover
from the drape window below the objective lens.
• The main plug shall only be connected to a socket with a faultless protective ground conductor.
BEFORE EVERY USE
• Always use the device within the scope of its intended use.
• When adding accessories and/or components please ensure the permissible total weight of
the device is not exceeded (Refer label "Maximum load")
• Adjust gas cylinder torque to compensate for any added weight if any; this will enable the
surgical microscope to maintain its balance in all positions of the working range.
• Before every use, limit the stroke of the suspension arm to prevent any contact with the
patient if the surgical microscope is lowered accidentally.
• Position the instrument in such a way that you can move it away from the patient at any time.
• When connecting accessories, make sure that sufficient free space is provided and that the
surgical microscope does not touch the patient.
• Before using a fundus viewing system, make sure that the room for movement is larger than
the downward travel path of the microscope.
• To prevent accessories from falling down, check before every use that accessories are
securely locked in position and securing screws are firmly tightened.
• Check that the movement of the surgical microscope is unobstructed to avoid any collisions
and damage to cables.
• When mounting non-Labomed equipment, make sure that sufficient free space is available
for focus positioning.
6219100-795
STELLA
Stella
8
Issue 1.0
July,2021