Intraoperative battery change
Risk of injury from hot bat-
tery!
Battery may be hot after use in
WARNING
the machine.
The battery removal device is used for intraoperative
battery changes ensuring that sterile conditions are
safely maintained.
Turn drill and reamer 6 so that the battery com-
partment points upward and remove cover 11.
Mount sterile battery removal device 18, see Fig. 7.
18
6
Fig. 7
Slightly shake drill and reamer 6 with battery
removal device 18 mounted and the battery com-
partment held pointing downward.
The battery slides smoothly into battery removal
device 18.
Hand over battery removal device 18 containing
the discharged battery to an unsterile person.
Insert a charged battery, see Inserting the battery.
Remove the battery with
the battery removal device
and let it cool.
8
Coupling and uncoupling of attachments on
the drill and reamer
Secure drill and reamer 6 against inadvertent acti-
vation with the safety catch 9.
Note
For further information about the attachments for the
drill and reamer, see TA011936 (Acculan 3Ti) and/or
TA014540 (Acculan 4).
To apply drilling wires we recommend using the special
Kirschner wire chuck. This quick-action chuck adapter
allows quick and easy tensioning of drilling wires.
To protect against injuries when using long drilling
wires, screw in the Kirschner wire protection sleeve
supplied with the product.
Screw Kirschner wire protection sleeve 15 into
adapter 17, see Fig. 8.
Fig. 8
17
15
en
11