7.1.3.
Solving Typical Problems
Error description
Display shows incorrect values
Long response time
High humidity values -
red LED blinks
Tab. 21 Self diagnosis
7.2
Cleaning the Sensing Head and Filter Replacement
In case of dusty, oily and polluted environment
▪
The filter cap shall be replaced once in a while with an E+E original one. A polluted filter cap causes longer
response time of the device.
▪
If needed, the sensing head can be cleaned. For cleaning instructions please see www.epluse.com/ee310.
7.3
RH and T Calibration and Adjustment
Humidity calibration and adjustment
Depending on the application and the requirements of certain industries, there might arise the need for periodical
humidity calibration (comparison with a reference) or adjustment (bringing the device in line with a reference).
Calibration and adjustment at E+E Elektronik
Calibration and/or adjustment can be performed in the E+E Elektronik calibration laboratory. For information on
the E+E capabilities in ISO or accredited calibration please see www.eplusecal.com.
Calibration and adjustment by the user
Depending on the level of accuracy required, the humidity reference can be:
▪
Humidity Calibrator (e.g. Humor 20), please see www.epluse.com/humor20.
▪
Handheld Device (e.g. Omniport30), please see www.epluse.com/omniport30.
▪
Humidity Calibration Kit (e.g. E+E Humidity Standards), please see www.epluse.com/ee310.
Perform 1 or 2 point adjustment via PCS10 Product Configuration Software or via display (see below).
Menu
Customer
adjustment
Rel. humidity
Temperature
Fig. 36
Adjustment menu
Likely causes and solutions
Error during re-adjustment of the sensor.
→ Reset to factory calibration and repeat the adjustment routine.
Filter polluted.
→ Replace filter
Output configured incorrectly
→ Check configuration
Filter polluted
→ Replace filter
Inappropriate filter type
→ Contact E+E representative for advice.
Water in sensing head
→ I nvestigate the cause for water / condensation.
Contact E+E representative for advice.
Inappropriate filter type
→ Contact E+E representative for advice.
Reference value
Offset
Current
measured value
Reference
value low
2 point adjustment
Reference value
high
i
Calibration status
Reset to factory calibration
value
Reference value
value
Current measured
value
Humidity and Temperature Sensor | 34
User Manual EE310