Metal sealing ring
(contained with the
probe as standard)
Extension
(not available for
NPT thread)
Fig. 11
Installation of the probe by utilizing the ball valve set
Removing the probe
1.
Hold the probe firmly to avoid it being suddenly expulsed when releasing the lock. Do not bend damage the
probe cable.
2.
Slowly loosen the lock nut with a spanner (spanner width 24) only till the probe is pushed out by the
overpressure in the process. Do not completely loosen the lock nut, but only as much as necessary for the
probe to slide.
3.
After the probe has been pushed out of the process up to the stop, close the ball valve.
4.
Remove the probe from the ball valve.
NOTICE
Observe the correct positioning of the sealing element 1 before reinstalling the probe.
Metal sealing ring
(as a standard contained
with the ball valve set)
1/2" ball valve
Adapter body
Inside diameter
≥13.1 mm (0.55")
Lock nut
User Manual EE310
Clamping sleeve
Sealing element 2
Sealing element 1
Humidity and Temperature Sensor | 17