Mounting with immersion well
The innovative stainless steel immersion well is mainly dedicated for measurement in liquids and allows for fast
and safe replacement of the T24 T only, remote probe.
Fig. 13
Immersion well for EE310-T24 T only, remote probe
Immersion Well Data
Pressure rating
Max. flow speed
Tab. 6
Data for immersion well
22
width across flats:
(0.87)
Ø8
(0.31)
Fig. 14
Immersion well dimensions in mm (inch)
The immersion well can also be used with a cable gland instead of the mounting spring.
Fig. 15
Handling of the immersion well
Procedure:
1.
The spring inside the well must be removed and replaced by a standard M12x1.5 cable gland
(not included in the scope of supply).
2.
Insert the sensor and fix it by fastening the cable gland.
NOTICE
Observe the operating temperature range of the cable gland to match the process parameters.
Innovative mounting spring
▪
For securing the probe inside the well
▪
No fastening screw, no tools required
25 bar (363 psi), stainless steel
9 m/s (1 771 ft/min)
1.
User Manual EE310
2.
Humidity and Temperature Sensor | 19