Functional Elements
The IEEE 488.2 standard requires a list of the functional elements which are
used by the MB100 receiver . These are the functional elements used in
constructing the remote commands that control the receiver. For more
information , see the IEEE 488.2 standard , sections 4.3 , 7.1.1 , and 7.3.3 . From
Tables 4.2 and 4.3 of the IEEE 488.2 standard , the receiver does the following:
< PROGRAM MESSAGE>
< PROGRAM MESSAGE UNIT >
<COMMAND MESSAGE UNIT>
< PROGRAM MESSAGE TERMINATOR >
< PROGRAM MESSAGE UNIT SEPARATOR >
<QUERY MESSAGE UNIT>
< COMMAND PROGRAM HEADER >*
< QUERY PROGRAM HEADER >*
< PROGRAM DATA SEPARATOR >
< PROGRAM HEADER SEPARATOR >
< PROGRAM DATA>
< DECIMAL NUMERIC PROGRAM DATA>
< NON- DECIMAL NUMERIC PROGRAM DATA>
< CHARACTER PROGRAM DATA>
<compound command program header> and < compound query program
header> are not handled .
*
Getting Started
Specific Command Implementations
The reset command " *rst" performs a device reset . As defined in the IEEE
488.2 , it will:
reset the device settings to default settings
Macros are not implemented in the MB100 , thus macros are ignored
force the receiver into Operation Complete Command Idle State ( OCIS) and
Operation Complete Query Idle State ( OQIS)
Self Test Query
This tests the receiver's basic functionality . The scope of the self test function is
limited .
Overlapped vs. Sequential Commands
All commands are sequential commands.
MB100 Programmer Manual
Operation Complete Message
All command actions are immediate (no overlapped commands ), such that
operation complete is immediate.