Skytron 1602 ESSENTIA Owner's Manual

Skytron 1602 ESSENTIA Owner's Manual

Surgical table

Advertisement

Quick Links

1602 ESSENTIA
SURGICAL TABLE
OWNER'S MANUAL
(Includes Operation, Maintenance and Parts)
Read this manual before operating the table! This information is necessary
for the safe and efficient operation of the equipment.
TEC-D-0108 REV5
09/27/2017

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 1602 ESSENTIA and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Skytron 1602 ESSENTIA

  • Page 1 1602 ESSENTIA SURGICAL TABLE OWNER'S MANUAL (Includes Operation, Maintenance and Parts) Read this manual before operating the table! This information is necessary for the safe and efficient operation of the equipment. TEC-D-0108 REV5 09/27/2017...
  • Page 2 The base language for this document is ENGLISH. Any translations must be from the base language document. Printed copies are not controlled documents. Although current at the time of publication, SKYTRON'S policy of continuous development makes this manual subject to change without notice. If current manuals are required, contact your local SKYTRON representative or contact SKYTRON...
  • Page 3: Table Of Contents

    3-9. Positioning Functions ......................15 3-10. Emergency Brake Release ....................18 3-11. Table Top Rotation ......................19 3-12. Head Section ........................20 3-13. Pad Sets ..........................21 3-14. Positioning ......................... 21 3-15. Positioning and Clearances ....................24 1602 ESSENTIA Owner's • REV5...
  • Page 4 4-9. Storage ..........................31 SECTION 5. REPLACEMENT PARTS ..................32 5-1. Head & Top Section Components ..................32 5-2. Base Components ......................34 SECTION 6. ELECTROMAGNETIC EMISSIONS ..............35 SECTION 7. REVISION HISTORY ..................... 39 1602 ESSENTIA Owner's • REV5...
  • Page 5: Section 1. Safety Information

    Prior to use, all personnel that may operate this table must be instructed in the correct operational a. Remove the power cord. procedures. The 1602 ESSENTIA surgical table is b. Place the main power switch in the OFF position. designed for use by trained and qualified personnel c.
  • Page 6 Page 4 WARNING Injury Hazard, Possible Pinch/Crush Points LEG SECTION CAN COLLIDE HEAD AND BACK SECTION WITH BASE, COLUMN OR FLOOR CAN COLLIDE WITH BASE OR FLOOR Figure 1. Possible Pinch/Crush Points 1602 ESSENTIA Owner's • REV5...
  • Page 7: Safety Precautions

    HEAD PATIENT’S approved replacement. SECTION HEAD Personal injury to patient or staff may result from a lack of proper BACK SECTION maintenance of this equipment. Figure 11. Operator's Position 1602 ESSENTIA Owner's • REV5...
  • Page 8 If circumstances demand table DO NOT use the table if any of the brakes be unlocked: inspection points fail. • The patient must be centered and evenly distributed on the table top (i.e. 1602 ESSENTIA Owner's • REV5...
  • Page 9 68°F (20° C). Batteries should be inoperable. re-charged every 6 months of product storage. Refer all service issues to a SKYTRON authorized service technician. Any parts or assemblies not listed in this section must be serviced or replaced CAUTION by SKYTRON authorized service personnel only.
  • Page 10: Section 2. Equipment Specifications

    Table has been wiped down to remove rust AIR, OXYGEN, or NITROUS OXIDE. inhibitors. • This product is not intended for sterilization. Items found to be non-conforming must be addressed prior to placing the table into service. 1602 ESSENTIA Owner's • REV5...
  • Page 11: Electrical Specifications

    (19”) (22-13/16”) (25-9/16”) ” 500mm (19-5/8”) TOP VIEW 60° 555mm (21-7/8”) 90° MAX. 1040mm (40-15/16”) MIN. 690mm (27-3/16”) 216mm (8-1/2”) 142mm (5-5/8”) 999mm 480mm (39-3/8”) (18-15/16”) SIDE VIEW FRONT VIEW Figure 2. 1602 Table Dimensions 1602 ESSENTIA Owner's • REV5...
  • Page 12: Equipment Labels

    INSTABILITÉ. FAIRE PREUVE DE VIGILANCE LORSOUE AN IMPROPERLY DISTRIBUTED PATIENT LOAD MAY LE PLATEAU DE LA TABLE TOURNE Á 90° CAR UNE CAUSE THE TABLE TO BE TIPPED OVER. CHARGE DE PATIENT NON RÉPARTIE UNIFORMÉMENT POURRAIT FAIRE BASCULER LA TABLE. 1602 ESSENTIA Owner's • REV5...
  • Page 13: Label Symbols

    Catalogue number ● ISO 15223-3.15 Serial number ● ISO 15223-3.16 AUTHORIZED REPRESENTATIVE IN ● ● EN980-5.13 THE EUROPEAN COMMUNITY Manufacturer ● ● EN980-5.12 General mandatory action sign ● ISO 7010-M001 Protective earth (ground) ● IEC 60417-5019 1602 ESSENTIA Owner's • REV5...
  • Page 14: Section 3. Operation

    Figure 3. 1602 Surgical Table 3-1. General The table is equipped with a removable 10.9 foot SKYTRON’s 1602 ESSENTIA Surgical Table is (3.3 m) long power cord with an approved hospital an electro-hydraulic operated, general purpose grade plug. The AC power switch (ON/OFF) is surgical table (Figure 3).
  • Page 15: Pendant Control

    Prior to operating the table, observe FLEX all table precaution labels and review Section 1-1. Special User Attention, in the front of this manual. BRAKE LOCK BRAKE UNLOCK LEVEL POWER ON/OFF Figure 6. Function Buttons 1602 ESSENTIA Owner's • REV5...
  • Page 16: Ac Operation

    (Refer to Figure 8). The battery indicator light, located in the upper part of the pendant control, will illuminate (Refer to Figure 9). The table is now ready for battery operation. 1602 ESSENTIA Owner's • REV5...
  • Page 17: Charging The Batteries

    With an evenly distributed patient weight system will automatically reduce input amperage load up to 500 lbs (227 kg), all table as capacity reaches 90% to avoid over charging. functions will operate smoothly and quietly. 1602 ESSENTIA Owner's • REV5...
  • Page 18 28° Re-lock the brakes immediately once the final position is reached and before commencing surgery. Press the BRAKE LOCK button on the pendant control to activate the brakes (Figure 12). Figure 13. Trendelenburg Positioning 1602 ESSENTIA Owner's • REV5...
  • Page 19 The back section will raise up to 90° above horizontal (Figure 15). Figure 16. Elevation Function To lower the table top, press the TABLE DOWN button. The table top will go down to a minimum height of 26" (660 mm). 1602 ESSENTIA Owner's • REV5...
  • Page 20: Emergency Brake Release

    FLEX button (Figure 18). REFLEX FLEX REFLEX REFLEX FLEX FLEX EMERGENCY BRAKE RELEASE Figure 20. Emergency Brake Release Figure 18. Flex / Reflex Positioning To return the table top to a horizontal position or into a reflex position, press the REFLEX button. 1602 ESSENTIA Owner's • REV5...
  • Page 21 Always lock the table top securely after rotating. activating any hydraulic function. CAUTION DO NOT rotate the table top less than 180°. There is a risk of injury to the patient and staff if the table were to tip over. 1602 ESSENTIA Owner's • REV5...
  • Page 22: Head Section

    ONLY. HEAD SECTION Two (2) locking knobs are located inside the leg LOCKING KNOBS section to secure the head section (Refer to Figure 24). Figure 23. Head Section Adjustment 1602 ESSENTIA Owner's • REV5...
  • Page 23: Pad Sets

    Compliance with IEC60601-1 edition 3 has been confirmed without the pad set. The SKYTRON pad sets are available in 2 inch (51 mm) or larger thickness and have a velcro strip which holds them in place on the table surface.
  • Page 24 1602 ESSENTIA General Purpose Patient Positioning Guidelines Upper Body Gall Bladder / Thoracic Vascular / Endovascular Kidney / Thoracic Lower Extremity / Edovascular / Podiatry Accessories shown may not be available in all markets. Contact your SKYTRON Representative for details. 1602 ESSENTIA Owner's • REV5...
  • Page 25 Page 23 1602 ESSENTIA General Purpose Patient Positioning Guidelines (Cont'd) Neurosurgery Shoulder Arthoroscopy Spinal Bariatric Surgery Accessories shown may not be available in all markets. Contact your SKYTRON Representative for details. 1602 ESSENTIA Owner's • REV5...
  • Page 26: Positioning And Clearances

    DO NOT REVERSE A PATIENT ON THIS TABLE CONSULT WITH SKYTRON FOR QUESIONS BACK FOOT END HEAD END OF BASE OF BASE CAUTION DO NOT reverse the patient on the table without first consulting with your SKYTRON representative. 1602 ESSENTIA Owner's • REV5...
  • Page 27: Section 4. Maintenance

    5. Use a ready-to-use detergent diluted as required by manufacturer. Ensure that the active WARNING ingredients of the product are compatible with the materials of the SKYTRON table. Personal injury to patient or staff may result from a lack of proper 6. Apply cleaning fluid liberally to top and sides maintenance of this equipment.
  • Page 28: Routine Inspections

    4. Inspect the power cord and plug for any c. Pressure check (with a gauge) the pressure signs of burns or damage. relief valve. 5. Test operation of the base power switch. 1602 ESSENTIA Owner's • REV5...
  • Page 29: Operator Troubleshooting

    Solenoid valve check valve leakage Contact SKYTRON Service Defective connecting gear due overweight operation Arm board cannot be adjusted Arm board was bent by siderail of Replace arm board or is binding back or seat section 1602 ESSENTIA Owner's • REV5...
  • Page 30: Maintenance Checks & Services

    5. Ensure the leg section is not searching (continual raising and lowering) when returning Refer all service issues to a SKYTRON to level. If needed, this adjustment can only be authorized service technician. performed by a SKYTRON authorized service technician.
  • Page 31 8. Secure the service access cover with the the notches inside the cap and press down four (4) screws removed in step 1. firmly. 6. Secure the service access cover with the four (4) screws removed in step 1. 1602 ESSENTIA Owner's • REV5...
  • Page 32: Maintenance Matrix

    The specific items listed in the MAINTENANCE optimal usage, safety and longevity of the product, MATRIX should be inspected and repaired or have it serviced only by a SKYTRON authorized replaced as necessary. The suggested time service technician using SKYTRON authorized intervals are intended as a guideline only and actual replacement parts and service techniques.
  • Page 33: Service

    Condition of hydraulic floor lock brake cylinders & pads 4-8. Disposal Instructions • Stability of table top The end of the useful life for the SKYTRON surgical table is 10 years under normal operating conditions. • Stability while table floor lock brakes are Service parts are available for this period. activated Contact your SKYTRON representative for disposal •...
  • Page 34: Section 5. Replacement Parts

    SECTION 5. REPLACEMENT PARTS CAUTION Replacement parts listed in this section have been identified by SKYTRON as servicable by facility Any parts or assemblies not listed in this personnel and are available for purchase. To obtain section must be serviced or replaced...
  • Page 35 (includes 22 hex nuts, 24 lock washers, 24 sleeves, 22 hex screws) D3-016-27 SIDE RAIL, back section, left ..................... 1 D3-016-30 SIDE RAIL, head section ....................1 (includes 22 hex nuts, 24 lock washers, 24 sleeves, 22 hex screws) D3-016-19 FRAME ASSEMBLY, head section ..................1 1602 ESSENTIA Owner's • REV5...
  • Page 36: Base Components

    POWER CORD ASSY (includes 1 outlet, 1 cable, 2 screws, 1 cable set, 1 fuse) ..... 1 D5-016-03 BATTERY, 12V (includes 2 batteries 4 wire caps, 1 bracket, 1 hex bolt, 1 washer) ....1 D1-016-03 COVER, service access (includes 4 truss head screws) ........... 1 1602 ESSENTIA Owner's • REV5...
  • Page 37: Section 6. Electromagnetic Emissions

    The 1602 Surgical Table is suitable for use in all Harmonic emissions establishments, other than domestic establishments Class A IEC 61000-3-2 and those directly connected to the public low-voltage power supply network that supplies buildings used for Voltage fluctuations / domestic purposes. flicker emissions Complies IEC 61000-3-3 1602 ESSENTIA Owner's • REV5...
  • Page 38 (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. 1602 ESSENTIA Owner's • REV5...
  • Page 39 3 A / m 3 A / m magnetic field location in a typical IEC 61000-4-8 commercial or hospital environment NOTE: U is the a.c. mains voltage prior to application of the test level. 1602 ESSENTIA Owner's • REV5...
  • Page 40 RF compliance level above, the 1602 Surgical Table should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re- orienting or relocating the 1602 Surgical Table. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 1602 ESSENTIA Owner's • REV5...
  • Page 41: Section 7. Revision History

    03/23/2015 Initial release. 05/20/2015 Added warning "Consult with Skytron before reversing a patient on the table" 07/02/2015 Changed part number on Pg 9 from J090X39 to D5-016-03. Fixed formatting issues on Page 9. Removed images of improper positioning on pages 22 and 23.
  • Page 42 Page 40 NOTES 1602 ESSENTIA Owner's • REV5...
  • Page 43 Page 41 NOTES 1602 ESSENTIA Owner's • REV5...
  • Page 44 5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI 49512 • 616.656.2900...