Page 1
SURGICAL TABLE OPERATORS MANUAL MODEL 1700...
Page 3
SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc. (SKYTRON) warrants all new products sold by it directly or through a dealer or other authorized representative, with exception to replacement parts, spares, bulbs (surgical lights), pads, and accessory items (surgical tables) to be free from defects in material or workmanship, under normal use and service, for a period of two (2) years.
TABLE OF CONTENTS Title Page EQUIPMENT LABELS AND SPECIFICATIONS ..................2 1700 Surgical Table Specifications ......................3 SPECIAL USER ATTENTION ........................ 4 SECTION I INTRODUCTION ......................... 7 1-1. General ............................7 1-2. Power Requirements ........................7 1-3. Pendant Control Unit ........................8 1-4.
EQUIPMENT LABELS AND SPECIFICATIONS INDICATES DANGEROUS VOLTAGE, 120 V, 60 Hz CLASS I DEFIBRILLATION PROOF, TYPE B EQUIPMENT- IPX4 RATED. INTERNALLY POWERED EQUIPMENT TYPE B EQUIPMENT PROTECTIVE GROUNDING. IN ORDER TO ENSURE PROPER GROUNDING RELIABILITY, THIS TABLE MUST BE CONNECTED TO A PROPERLY GROUNDED HOSPITAL GRADE OUTLET.
SPECIAL USER ATTENTION The positioning capabilities of the 1700 Table requires special attention for possible interference points when using multiple function positioning. As with the operation of any surgical table, a certain amount of care should be exercised to position the patient safely.
Page 9
SPECIAL USER ATTENTION WARNING WARNING DO NOT unlock brakes when a patient Risk of electrical shock. Make sure power cord is disconnected prior to is on the table. An uneven patient accessing fuses. weight load may cause instability. NOTE WARNING Activating any function button will acti- vate the brake system.
Page 10
DO NOT use cleaners con- taining phenolics. WARNING Certain accessories may limit weight CAUTION capacities, check with your SKYTRON representative. When using spray cleaners DO NOT spray fluids directly into electrical re- ceptacles. NOTE...
SWITCH EMERGENCY BRAKE RELEASE Figure 1-1. Elite 1700 1-1. General SKYTRON’s Elite 1700 Surgical Table is an electro- hydraulically operated, general purpose surgical table. See figure 1-1. The electro-hydraulic positioning functions oper- ated by the hand-held, push button, pendant control...
1-3. Pendant Control Unit 1-4. Floor Lock/Brake System The hand-held pendant control unit (figure 1-3) has The floor lock/brake system consists of four self- a spring clip hanger located on the back of the leveling, hydraulic brake cylinders which raise and control for storage.
SECTION II OPERATION 2-2. Positioning Functions 2-1. Electrical Power The hand-held pendant control (figure 2-2) acti- a. Check to be sure the power cord is plugged vates the following table functions: into a properly grounded, Hospital Grade, 120VAC outlet. Make sure the power cord is routed so as to prevent it from being in the way of the operating personnel.
a. Floor Lock/Brake System. To activate the b. Trendelenburg. To place the surgical table in a Trendelenburg (head down) position, press the brakes without affecting table positioning, press the TREND button (figure 2-4). Trendelenburg position- TABLE UP button. See figure 2-3. The elevation ing of up to 25°...
LATERAL LATERAL TILT RIGHT TILT LEFT TABLE TABLE DOWN 25˚ 25˚ 39" 29" Figure 2-7. Elevation Function Figure 2-5. Lateral Tilt Positioning f. Leg Section. To lower the leg section press the LEG DOWN button (figure 2-8). The leg section will go down to 90°...
UNLOCK button pulling it straight out of the back section. 1700 Tables on the pendant control will have to be have the capability of attaching the head section to the pressed before brakes will lock again.
2-5. 180 Degree Table Top Rotation. NOTE WARNING Normal table top position is with the head (and back) section over the power Exercise caution with the table top ro- cord end of the base. tated 90° to the base since an improp- erly distributed patient load may cause a.
2-6. Positioning The use of certain optional accessories available from SKYTRON further extend the positioning capabilities of the 1700 Table. Refer to the following Positioning Guidelines or contact your SKYTRON representative for further details. WARNING Certain accessories may limit weight capacities.
NOTE The following preventive maintenance checks and services are recommended to ensure the service- Always follow current AORN Journal ability and proper operation of your SKYTRON Guidelines to ensure proper cleaning Surgical Table. and disinfection procedure. a. During normal cleaning, a general visual...
To obtain service instructions, replacement parts, factory service or preventive maintenance con- Repeat cleaning procedure for all table surfaces including the top, sides, elevation column, base tracts, contact your nearest SKYTRON represen- and all accessories. tative or write or call: CAUTION...
Page 24
5000 36th Street S.E., Grand Rapids, MI 49512 1-800-SKYTRON or 1-616-957-0500 • FAX 1-616-957-5053...