Service - Endress+Hauser Liquisys M COM 223 Operating Instructions Manual

Transmitter for dissolved oxygen
Hide thumbs Also See for Liquisys M COM 223:
Table of Contents

Advertisement

Instrument configuration
Coding
Field
Function group
S

SERVICE

Selection of
S1
language
S2
Configure hold
S3
Manual hold
Enter hold dwell
S4
period
Enter release
code for software
S5
upgrade on Plus
package
Enter release
code for software
S6
upgrade on
ChemoClean
Order code is
S7
displayed
50
5.6
Service
Selection or range
Factory setting
ENG = English
GER = German
FRA = French
ITA = Italian
NEL = Dutch
ESP = Spanish
S+C = During
configuration
and calibration
CAL = During
calibration
Setup = During
configuration
none = No hold
Off
On
10 s
0 ... 999 s
0
0 ... 9999
0
0 ... 9999
Liquisys M COM 223 / 253
Display
S
SERVICE
This field must be configured
ENG
once during start-up. Afterwards,
S1
you can leave S1 and continue
Language
(see Chapter 5.1).
S+C
S = Setup,
S2
C = Calibration.
Auto HOLD
Off
The setting remains active even
S3
after a power failure.
Man. HOLD
s
10
S4
Cont.Time
When the wrong code is entered,
the system jumps back to the
measurement menu. The number
0
is edited with the PLUS or MINUS
S5
key and confirmed with the
PlusCode
ENTER key.
"1" is displayed in case of active
code.
When the wrong code is entered,
the system jumps back to the
measurement menu. The number
0
is edited with the PLUS or MINUS
S6
key and confirmed with the
CleanCode
ENTER key.
"1" is displayed in case of active
code.
When you upgrade the device,
order
the order code is updated
S7
automatically. The delivery state
DX0005
will be displayed.
Info
Endress+Hauser