Instrument configuration
Coding
Field
Enter
K6
uncompensated
conductivity value
Enter
K7
concentration
value for K6
Enter temperature
K8
value for K6
Enter whether or
K9
not the table
status is ok
Coding
Field
Function group
S

SERVICE

Selection of
S1
language
S2
Hold configuration
Factory settings are printed in bold face;
base version does not include functions in italic .
46
Selection or range
Factory setting
0.0 µS/cm
0.0 ... 9999 mS/cm
0.00 %
0.00 ... 99.99 %
0.0 °C
–35.0 ... 250.0 °C
yes
no
5.8
Service
Selection or range
Factory setting
ENG = English
GER = German
FRA = French
ITA = Italian
NEL = Dutch
ESP = Spanish
S+C = during
setup and
calibration
CAL = during
calibration
Setup = during
setup
No
= no hold
Liquisys M CLM 223 / 253
Display
The function chain K5 ... K8 will
run through automatically as
0.0
µS/cm
K6
many times as corresponds to
the value in K4. Then the system
conduct.
jumps to K9.
0.0
%
K7
Measuring unit selected as in A2.
concentr.
°C
0.0
K8
Temp.val.
Return to K.
yes
If not, then set table correctly (all
K9
previous settings are kept) or
Status ok
back to measurement mode (this
makes the table invalid).
Display
S
SERVICE
This field must be set configured
ENG
once during device start-up.
S1
After confirmation with ENTER
Language
you can exit S1 and continue.
S+C
S = setup,
S2
C = calibration.
Auto HOLD
Info
Info
Endress+Hauser