Changing The Filter; Cleaning And Maintenance - Beurer IH 24 Kids Instructions For Use Manual

Hide thumbs Also See for IH 24 Kids:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 16
Important
• Only use the nasal douche with isotonic saline
solutions.
• Replace the nasal douche after a year.
The maximum volume of the nasal douche is 10 ml.
If you are unsure about the correct dilution of the
saline solution, please consult your pharmacist.
7.9 Using the nasal douche
To use the nasal douche correctly, proceed as fol-
lows
:
B
1. Twist the cover
17
move it in an upward direction.
2. Also lift the collection container
3. Fill the saline container
4. Place the collection container
container
.
19
5. Place the cover
20
Twist the cover
20
it clicks firmly into place.
6. Connect the compressed air hose
line container
19
7. Carefully position the cover
nostrils. Breathe through your nose.
8. Press the On/Off control
cation, place one finger over the cover hole
on the container.
Notice
During the application, breathe slowly and evenly
through your nose. Tilt your head slightly in the op-
posite direction to the nostril being treated so that
the saline solution can flow deep into the nose.
9. To pause or stop the application, take your finger
off the cover hole
20
10. Once your nose feels clear again, you can
stop the application. The application should
be stopped at the latest once the maximum
filling quantity has been used.
7.10 Cleaning
See "Cleaning and maintenance".

8. CHANGING THE FILTER

In normal operating conditions, the air filter must be
replaced after approx. 200 operating hours or one
year. Please check the air filter regularly (after 10-
12 nebulisation procedures). Replace the used filter
if it is very dirty or clogged. If the filter has become
damp, it must also be exchanged for a new filter.
90° anticlockwise and re-
out.
18
with saline solution.
19
into the saline
18
on the saline container
in a clockwise direction until
to the sa-
7
from below.
in one of your
20
. To start the appli-
2
on the saline container
Important
• Do not attempt to clean the used filter and re-
use it.
• Only use the manufacturer's original filter, oth-
erwise your nebuliser may become damaged
and sufficiently effective treatment cannot be
guaranteed.
• Do not repair or maintain the air filter while it
is in use.
• Never operate the device without a filter.
To replace the filter, proceed as follows:
Important
• First switch the device off and disconnect it
from the mains.
• Allow the device to cool down.
.
19
1. Remove the filter cap
Notice
If the filter remains in the device after the cap has
been removed, take the filter out of the device, e.g.
with tweezers or similar.
2. Re-insert the filter cap
3. Ensure that it is securely in place.
20
9. CLEANING AND MAINTE-
NANCE
Atomiser and atomiser set
Warning
Adhere to the following hygiene instructions to
.
19
avoid health risks.
• The atomiser
set are intended for multiple use. Please note
that different areas of application involve dif-
ferent requirements in terms of cleaning and
hygienic preparation.
Notice
• Do not clean the atomiser or the atomiser set
mechanically using a brush or similar device,
as this could cause irreparable damage and it
will mean that the best treatment results can no
longer be guaranteed.
• Please consult your doctor about the addition-
al requirements in terms of the hygienic prepa-
ration needed (hand care, handling of medica-
22
upwards.
5
C
with a new filter.
5
and the rest of the atomiser
8